Clinical Investigation
Comparison of aortic annular diameter defined by different measurement mordalities before transcatheter aortic valve implantation
Ruixiang Qi, Xiangdong You, Zhaoxia Pu, Qian Yang, Zhaoxu Huang, Liming Zhou, Pintong Huang
Published 2017-05-24
Cite as Chin J Cardiol, 2017, 45(5): 409-414. DOI: 10.3760/cma.j.issn.0253-3758.2017.05.009
Abstract
ObjectiveTo compare aortic annular diameter measured by transthoracic echocardiography (TTE), transesophageal echocardiography (TEE), and multislice computed tomography (MSCT) in patients with severe aortic stenosis, and to evaluate the impact on selection of prosthetic valve type in transcatheter aortic valve implantation (TAVI).
MethodsClinical data of 138 patients with severe aortic stenosis referred for TAVI between January 2014 and June 2016 in our hospital were retrospectively analyzed.The difference of aortic annular diameter measured by TTE, TEE, and MSCT were compared.TTE was performed after TAVI to evaluate the accuracy of measurement before TAVI.
Results(1) Aortic annular diameter was (23.37±2.22) mm by TTE and (23.52±1.70) mm by TEE (P=0.12). Pearson correlation analysis showed that aortic annular diameter measured by TTE was correlated to that measured by TEE (r=0.87, P<0.05). (2)The long-axis diameter and the short-axis diameter measured by MSCT multiplanar reconstruction were significantly different ((27.86±2.87) mm vs. (21.91±2.53) mm, P<0.05). There was no significant difference between the mean of the long- and short-axis diameters and the diameter derived from cross-sectional area by MSCT ((24.92±2.38) mm vs. (24.84±2.25) mm, P=0.08). However, the diameter derived from the circumference ((25.35±2.34) mm) was significantly larger than the mean of the long- and short-axis diameters and the diameter derived from cross-sectional area by MSCT, and the difference were (0.43±0.62) mm and (0.51±0.62) mm respectively(both P<0.05). (3) Measurements of the aortic annulus diameter by MSCT including the mean of the long- and short-axis diameters, diameter derived from cross-sectional area, and diameter derived from the circumference were larger than the measurement by TTE and TTE (all P<0.05). (4) Implantation was successful in all patients.Moderate to severe paravalvular leakage was detected in 3 patients at 7 days post TAVI, and 1 patient developed severe prosthetic valve restenosis at 6 months post TAVI and received valve-in-valve implantation.
ConclusionsIn severe aortic stenosis patients referred for TAVI, the aortic annular diameter derived from TTE and TEE measurements are smaller than that from MSCT.In the absence of a gold standard, selection of prosthetic valve type in TAVI procedure should rely on comprehensive considerations, which is of importance to get good clinical results for severe aortic stenosis patients underwent TAVI.
Key words:
Aortic valve stenosis; Echocardiography; Tomography, X-ray computed
Contributor Information
Ruixiang Qi
Heart Center, Second Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou 310009, China
Xiangdong You
Zhaoxia Pu
Qian Yang
Zhaoxu Huang
Liming Zhou
Pintong Huang