Comparison of IQ-SPECT and low energy high-resolution collimator SPECT in clinical application
Han Kai, Zhang Hailong, Zhang Zongyao, Sun Xiaoxin, Wang Lei, Fang Wei
Published 2022-06-25
Cite as Chin J Nucl Med Mol Imaging, 2022, 42(6): 357-362. DOI: 10.3760/cma.j.cn321828-20201117-00415
Abstract
ObjectiveTo investigate differences in cardiac function and perfusion parameters measured by IQ-SPECT and low energy high-resolution (LEHR)-SPECT, as well as effects of scattering correction (SC) and CT attenuation correction (AC) on myocardial perfusion imaging.
MethodsFrom May 2020 to September 2020, 80 patients (58 males, 22 females, age (57±10) years) who underwent SPECT myocardial perfusion imaging were retrospectively enrolled in Fuwai Hospital. According to the standardized left ventricular end-diastolic diameter of body surface measured by two-dimensional echocardiography, patients were divided into 2 groups: A group (n=34) with significantly enlarged left ventricle and B group (n=46) with no significant enlargement of left ventricle. LEHR-SPECT and IQ-SPECT gated myocardial perfusion imaging were performed in all patients. Left ventricular end-diastolic volume (EDV), end-systolic volume (ESV), ejection fraction (EF), summed rest score (SRS) and total perfusion defect (TPD) were measured. Parameters measured by two methods and differences before and after SC and AC were compared by using paired t test and Wilcoxon signed rank test. The correlation was analyzed by Pearson correlation or Spearman rank correlation analyses, and the consistency was analyzed by Bland-Altman analysis.
ResultsIn A group, EDV, EF and SRS measured by IQ-SPECT were significantly lower than those assessed by LEHR-SPECT after SC and AC (EDV: (257±137) vs (276±154) ml, EF: (21±11)% vs (26±13)%, SRS: 17(6, 25) vs 18(8, 28); t values: -2.63, -7.46, z=-2.14, all P<0.05); all parameters measured by LEHR-SPECT and IQ-SPECT had well correlation (r values: 0.965, 0.969, 0.967, rs values: 0.920, 0.960, all P<0.001) and consistency. In B group, EDV and EF measured by IQ-SPECT were significantly lower than those assessed by LEHR-SPECT after SC and AC (EDV: (96±40) vs (107±39) ml, EF: (46±15)% vs (54±16)%; t values: -6.23, -10.71, both P=0.001); SRS and TPD measured by IQ-SPECT after SC and AC were significantly lower than non-SC and non-AC (SRS: 2(1, 4) vs 5(3, 11), TPD: (3%(1%, 5%) vs 7%(3%, 12%); z values: -4.11, -4.16, both P<0.001); all parameters measured by LEHR-SPECT and IQ-SPECT had well correlation (r values: 0.956, 0.978, 0.958, rs values: 0.926, 0.944, all P<0.001) and consistency.
ConclusionsThere are good correlation and consistency of left ventricular function and myocardial perfusion parameters acquired by IQ-SPECT and LEHR-SPECT. Moreover, IQ-SPECT is able to shorten acquisition time, resulting in great potential in clinical application.
Key words:
Myocardial perfusion imaging; Tomography, emission-computed, single-photon; Technetium Tc 99m sestamibi; Image processing, computer-assisted
Contributor Information
Han Kai
Department of Nuclear Medicine, Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China
Zhang Hailong
Department of Nuclear Medicine, Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China
Zhang Zongyao
Department of Nuclear Medicine, Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China
Sun Xiaoxin
Department of Nuclear Medicine, Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China
Wang Lei
Department of Nuclear Medicine, Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China
Fang Wei
Department of Nuclear Medicine, Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China