IOLMaster 700与Pentacam AXL测量白内障患者眼生物学参数的一致性
李栋军
杨文利
王子杨
陈伟
赵琦
李逸丰
崔蕊
沈琳
鲜军舫
作者及单位信息
·
DOI: 10.3760/cma.j.cn115989-20200924-00663
Agreement of biometry parameters measured by IOLMaster 700 and Pentacam AXL in cataract eyes
Li Dongjun
Yang Wenli
Wang Ziyang
Chen Wei
Zhao Qi
Li Yifeng
Cui Rui
Shen Lin
Xian Junfang
Authors Info & Affiliations
Li Dongjun
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing 100730, China
Yang Wenli
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing 100730, China
Wang Ziyang
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing 100730, China
Chen Wei
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing 100730, China
Zhao Qi
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing 100730, China
Li Yifeng
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing 100730, China
Cui Rui
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing 100730, China
Shen Lin
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing 100730, China
Xian Junfang
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing 100730, China
·
DOI: 10.3760/cma.j.cn115989-20200924-00663
0
0
0
0
0
0
PDF下载
APP内阅读
摘要

目的比较IOLMaster 700与Pentacam AXL测量白内障患者眼生物学参数的一致性。

方法采用横断面研究方法,纳入2018年1月2日—11日于北京同仁医院眼科中心就诊的单纯白内障患者76例146眼,其中男36例69眼,女40例77眼;受检者年龄35~88岁,平均(64.52±11.36)岁。所有患者均分别采用IOLMaster 700及Pentacam AXL测量眼轴长度(AL)、平均角膜曲率(Km)、前房深度(ACD)、中央角膜厚度(CCT)和角膜直径(WTW)。比较2种检查仪器各测量值的差异,评估两种测量结果的相关性和一致性。

结果IOLMaster 700与Pentacam AXL测量的AL分别为(24.03±1.72)mm和(23.98±1.73)mm,Km分别为(44.04±1.71)D和(43.88±1.72)D,ACD分别为(3.24±0.47)mm和(3.28±0.47)mm,WTW分别为(11.82±0.51)mm和(11.49±0.49)mm,IOLMaster 700测得的AL、Km、WTW值较大,ACD值较小,差异均有统计学意义( t=18.06、8.94、23.83、-15.36,均 P<0.01),IOLMaster 700与Pentacam AXL测量的CCT分别为(534.18±30.12)μm和(533.42±28.71)μm,差异无统计学意义( t=1.04, P=0.30)。Pearson线性相关分析结果显示,2种仪器测量的AL、Km、ACD、CCT及WTW值均呈明显正相关( r=0.999、0.992、0.998、0.956、0.942,均 P<0.01)。Bland-Altman分析结果显示,2种仪器测量的AL、Km、ACD、CCT及WTW值的95%一致性界限分别为-0.02~0.11 mm、-0.27~0.59 D、-0.10~0.02 mm、-16.5~18.1 μm及0.00~0.67 mm。

结论IOLMaster 700与Pentacam AXL测量AL、ACD、CCT具有较好的一致性及较窄的95%一致性界限,临床上可以互换使用。两种仪器测量WTW和Km的95%一致性界限超出临床可接受范围,均不建议互换使用。

生物测量;扫描光相干断层扫描;Scheimpflug成像;眼轴长度;角膜曲率;前房深度;角膜厚度;角膜直径
ABSTRACT

ObjectiveTo compare the agreement of biometric parameters measured by a new swept-source optical coherence tomography IOLMaster 700 and a new Scheimpflug imaging combined with partial coherence interferometry biometer Pentacam AXL in cataractous eyes.

MethodsA cross-sectional study was performed.A total of 146 eyes from 76 cataract patients were enrolled in Beijing Tongren Eye Center from 2 to 11 in January 2018, including 69 eyes in 36 males and 77 eyes in 40 females, the age ranged from 35 to 88 years, the mean age was (64.52±11.36) years.Axial length (AL), mean keratometry (Km), anterior chamber depth (ACD), central corneal thickness (CCT) and white-to-white (WTW) were measured with IOLMaster 700 and Pentacam AXL, respectively.The differences and correlations of the biometry parameters between the two devices were assessed.Bland-Altman method was used to evaluate the agreement of these parameters between the two devices and the 95% limits of agreement was calculated.This study adhered to the tenets of the Declaration of Helsinki and was approved by an Ethics Committee of Beijing Tongren Hospital (TRECKY2018-049).

ResultsThe mean values of the AL measured by IOLMaster 700 and Pentacam AXL were (24.03±1.72)mm and (23.98±1.73)mm, Km were (44.04±1.71)D and (43.88±1.72)D, ACD were (3.24±0.47)mm and (3.28±0.47)mm, WTW were (11.82±0.51)mm and (11.49±0.49)mm, respectively.Compared with the values in Pentacam AXL, the AL, Km and WTW were larger while ACD was smaller in IOLMaster 700 with significant differences ( t=18.06, 8.94, 23.83, -15.36, all at P<0.01). The mean CCT values from the IOLMaster 700 and Pentacam AXL were (534.18±30.12)μm and (533.42±28.71)μm, respectively, with no significant difference ( t=1.04, P=0.30). The AL, Km, ACD, CCT and WTW from IOLMaster 700 were highly correlated with those from Pentacam AXL ( r=0.999, 0.992, 0.998, 0.956, 0.942; all at P<0.01). The Bland Altman analysis showed that the 95% limits of agreement of AL, Km, ACD, CCT and WTW were-0.02-0.11 mm, -0.27-0.59 D, -0.10-0.02 mm, -16.5-18.1 μm and 0.00-0.67 mm.

ConclusionsIOLMaster 700 and Pentacam AXL have good agreements and narrow 95% limits of agreement in the measurements of AL, ACD, CCT, which can be clinically interchangeable.The 95% limits of agreement of WTW and Km are out of the clinically acceptable range, so the WTW and Km measured by the two devices are not interchangeable.

Biometry;Swept-source optical coherence tomography;Scheimpflug image;Axial length;Mean keratometry;Anterior chamber depth;Central corneal thickness;White to white
Yang Wenli, Email: mocdef.3ab61rt_lwgnay
引用本文

李栋军,杨文利,王子杨,等. IOLMaster 700与Pentacam AXL测量白内障患者眼生物学参数的一致性[J]. 中华实验眼科杂志,2020,38(11):962-966.

DOI:10.3760/cma.j.cn115989-20200924-00663

PERMISSIONS

Request permissions for this article from CCC.

评价本文
*以上评分为匿名评价
屈光性白内障手术对于人工晶状体(intraocular lens,IOL)度数的计算提出了更高的要求,白内障术后屈光误差产生的原因主要来源于测量误差 [ 1 ],因此精准的生物测量是保证IOL度数精确计算、患者术后获得最佳视力的前提。光学生物测量仪由于具有更高的精确性和可重复性、操作简便以及对操作者技术要求低等特点,已经成为眼生物参数测量、IOL度数计算的主要方法 [ 2 , 3 , 4 ]。IOLMaster曾被认为是白内障术前生物测量的金标准,新一代IOLMaster 700在原有仪器的基础上进行了多项技术升级,而Pentacam AXL是在Pentacam基础上的第1代可进行白内障术前生物测量的仪器。目前,对于这2种测量仪测量眼生物参数比较的报道相对较少。本研究旨在比较2种新型仪器测量的眼生物测量值的一致性,为临床应用提供参考。
试读结束,您可以通过登录机构账户或个人账户后获取全文阅读权限。
参考文献
[1]
Jin GJ , Crandall AS , Jones JJ . Intraocular lens exchange due to incorrect lens power[J]Ophthalmology 2007114(3):417-424. DOI: 10.1016/j.ophtha.2006.07.041 .
返回引文位置Google Scholar
百度学术
万方数据
[2]
Lam AK , Chan R , Pang PC . The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster[J]Ophthalmic Physiol Opt 200121(6):477-483. DOI: 10.1046/j.1475-1313.2001.00611.x .
返回引文位置Google Scholar
百度学术
万方数据
[3]
Haigis W , Lege B , Miller N et al. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis[J]Graefe's Arch Clin Exp Ophthalmol 2000238(9):765-773. DOI: 10.1007/s004170000188 .
返回引文位置Google Scholar
百度学术
万方数据
[4]
Kolega , Kovačević S , Čanović S et al. Comparison of IOL—master and ultrasound biometry in preoperative intra ocular lens (IOL) power calculation[J]Coll Antropol 201539(1):233-235.
返回引文位置Google Scholar
百度学术
万方数据
[5]
范雯邢怡桥Pentacam三维眼前节分析诊断系统在眼前节影像分析中的应用进展[J]中华实验眼科杂志 201230(2):176-179. DOI: 10.3760/cma.j.issn.2095-0160.2012.02.021 .
返回引文位置Google Scholar
百度学术
万方数据
Fan W , Xing YQ . Current progress in the application of Pentacam scheimpflug system in anterior eye segment photography[J]Chin J Exp Ophthalmol 201230(2):176-179. DOI: 10.3760/cma.j.issn.2095-0160.2012.02.021 .
Goto CitationGoogle Scholar
Baidu Scholar
Wanfang Data
[6]
Olsen T Calculation of intraocular lens power:a review[J]Acta Ophthalmol Scand 200785(5):472-485. DOI: 10.1111/j.1600-0420.2007.00879.x .
返回引文位置Google Scholar
百度学术
万方数据
[7]
Cass K , Thompson CM , Tromans C et al. Evaluation of the validity and reliability of A-scan ultrasound biometry with a single use disposable cover[J]Br J Ophthalmol 200286(3):344-349. DOI: 10.1136/bjo.86.3.344 .
返回引文位置Google Scholar
百度学术
万方数据
[8]
Eleftheriadis H IOLMaster biometry:refractive results of 100 consecutive cases[J]Br J Ophthalmol 200387(8):960-963. DOI: 10.1136/bjo.87.8.960 .
返回引文位置Google Scholar
百度学术
万方数据
[9]
Sel S , Stange J , Kaiser D et al. Repeatability and agreement of Scheimpflug-based and swept-source optical biometry measurements[J]Cont Lens Anterior Eye 201740(5):318-322. DOI: 10.1016/j.clae.2017.03.007 .
返回引文位置Google Scholar
百度学术
万方数据
[10]
Shajari M , Cremonese C , Petermann K et al. Comparison of axial length,corneal curvature,and anterior chamber depth measurements of 2 recently introduced devices to a known biometer[J]Am J Ophthalmol 2017178:58-64. DOI: 10.1016/j.ajo.2017.02.027 .
返回引文位置Google Scholar
百度学术
万方数据
[11]
Özyol P , Özyol E Agreement between swept-source optical biometry and Scheimpflug-based topography measurements of anterior segment parameters[J]Am J Ophthalmol 2016169:73-78. DOI: 10.1016/j.ajo.2016.06.020 .
返回引文位置Google Scholar
百度学术
万方数据
[12]
Ruiz-Mesa R , Abengózar-Vela A , Ruiz-Santos M Comparison of a new Scheimpflug imaging combined with partial coherence interferometry biometer and a low-coherence reflectometry biometer[J]J Cataract Refract Surg 201743(11):1406-1412. DOI: 10.1016/j.jcrs.2017.08.016 .
返回引文位置Google Scholar
百度学术
万方数据
[13]
Olsen T Prediction of the effective postoperative (intraocular lens) anterior chamber depth[J]J Cataract Refract Surg 200632(3):419-424. DOI: 10.1016/j.jcrs.2005.12.139 .
返回引文位置Google Scholar
百度学术
万方数据
[14]
李岩成拾明杨欣Sirius和Pentacam测量前房深度及中央角膜厚度值的比较研究[J]中华实验眼科杂志 201230(3):262-265. DOI: 10.3760/cma.j.issn.2095-0160.2012.03.019 .
返回引文位置Google Scholar
百度学术
万方数据
Li Y , Cheng SM , Yang X et al. Comparison of anterior chamber depth and central corneal thickness measuring values between Sirius and Pentacam[J]Chin J Exp Ophthalmol 201230(3):262-265. DOI: 10.3760/cma.j.issn.2095-0160.2012.03.019 .
Goto CitationGoogle Scholar
Baidu Scholar
Wanfang Data
[15]
Doughty MJ , Zaman ML . Human corneal thickness and its impact on intraocular pressure measures:a review and meta-analysis approach[J]Surv Ophthalmol 200044(5):367-408. DOI: 10.1016/s0039-6257(00)00110-7 .
返回引文位置Google Scholar
百度学术
万方数据
[16]
Read SA , Collins MJ , Iskander DR . Diurnal variation of axial length,intraocular pressure,and anterior eye biometrics[J]Invest Ophthalmol Vis Sci 200849(7):2911-2918. DOI: 10.1167/iovs.08-1833 .
返回引文位置Google Scholar
百度学术
万方数据
[17]
成拾明李岩黄锦海Pentacam与A型超声测量正常人中央角膜厚度比较的Meta分析[J]中华实验眼科杂志 201331(2):172-176. DOI: 10.3760/cma.j.issn.2095-0160.2013.02.016 .
返回引文位置Google Scholar
百度学术
万方数据
Cheng SM , Li Y , Huang JH et al. Meta analysis of comparison of central corneal thickness measurement between Pentacam and A-scan in normal people[J]Chin J Exp Ophthalmol 201331(2):172-176. DOI: 10.3760/cma.j.issn.2095-0160.2013.02.016 .
Goto CitationGoogle Scholar
Baidu Scholar
Wanfang Data
[18]
Prager TC , Hardten DR , Fogal BJ . Enhancing intraocular lens outcome precision:an evaluation of axial length determinations,keratometry,and IOL formulas[J]Ophthalmol Clin North Am 200619(4):435-448. DOI: 10.1016/j.ohc.2006.07.009 .
返回引文位置Google Scholar
百度学术
万方数据
[19]
Gharaee H , Abrishami M , Shafiee M et al. White-to-white corneal diameter:normal values in healthy Iranian population obtained with the Orbscan II[J]Int J Ophthalmol 20147(2):309-312. DOI: 10.3980/j.issn.2222-3959.2014.02.20 .
返回引文位置Google Scholar
百度学术
万方数据
[20]
Salouti R , Nowroozzadeh MH , Zamani M et al. Comparison of horizontal corneal diameter measurements using the Orbscan IIz and Pentacam HR systems[J]Cornea 201332(11):1460-1464. DOI: 10.1097/ICO.0b013e3182a40786 .
返回引文位置Google Scholar
百度学术
万方数据
[21]
Ferrer-Blasco T , Esteve-Taboada JJ , Martínez-Albert N et al. Agreement of white-to-white measurements with the IOLMaster 700,Atlas 9000,and Sirius systems[J]Expert Rev Med Devices 201815(6):453-459. DOI: 10.1080/17434440.2018.1481745 .
返回引文位置Google Scholar
百度学术
万方数据
备注信息
A
杨文利,Email: mocdef.3ab61rt_lwgnay
B
所有作者均声明不存在任何利益冲突
C
北京市医院管理局临床医学发展专项经费项目 (ZYLX201704)
北京市卫生系统高层次人才培养计划学科带头人项目 (2014-2-005)
评论 (0条)
注册
登录
时间排序
暂无评论,发表第一条评论抢沙发
MedAI助手(体验版)
文档即答
智问智答
机器翻译
回答内容由人工智能生成,我社无法保证其准确性和完整性,该生成内容不代表我们的态度或观点,仅供参考。
生成快照
文献快照

你好,我可以帮助您更好的了解本文,请向我提问您关注的问题。

0/2000

《中华医学会杂志社用户协议》 | 《隐私政策》

《SparkDesk 用户协议》 | 《SparkDesk 隐私政策》

网信算备340104764864601230055号 | 网信算备340104726288401230013号

技术支持:

历史对话
本文全部
还没有聊天记录
设置
模式
纯净模式沉浸模式
字号