OPD-Scan Ⅲ测量角膜屈光力和散光的可重复性及其与Pentacam检测结果的一致性
李盼盼
袁幽
黄黎黎
朱妍
李俊杰
宋愈
作者及单位信息
·
DOI: 10.3760/cma.j.cn115989-20190323-00134
Repeatability of OPD-Scan Ⅲ and its consistency with Pentacam in measuring keratometry and astigmatism
Li Panpan
Yuan You
Huang Lili
Zhu Yan
Li Junjie
Song Yu
Authors Info & Affiliations
Li Panpan
Department of Ophthalmology, Second Affiliated Hospital of Nantong University, Nantong First People's Hospital, Nantong 226001, China
Yuan You
Department of Ophthalmology, Lixiang Eye Hospital of Soochow University, Suzhou 215000, China
Huang Lili
Department of Ophthalmology, Second Affiliated Hospital of Nantong University, Nantong First People's Hospital, Nantong 226001, China
Zhu Yan
Department of Ophthalmology, Second Affiliated Hospital of Nantong University, Nantong First People's Hospital, Nantong 226001, China
Li Junjie
Department of Ophthalmology, Second Affiliated Hospital of Nantong University, Nantong First People's Hospital, Nantong 226001, China
Song Yu
Department of Ophthalmology, Second Affiliated Hospital of Nantong University, Nantong First People's Hospital, Nantong 226001, China
·
DOI: 10.3760/cma.j.cn115989-20190323-00134
0
0
0
0
0
0
PDF下载
APP内阅读
摘要

目的分析OPD-Scan Ⅲ测量角膜屈光力及散光的可重复性及其与Pentacam检测结果的一致性。

方法采用诊断试验研究方法,收集2018年8月于南通大学第二附属医院及苏州大学附属理想眼科医院就诊的屈光不正患者50例100眼,采用主觉验光测量受检者球镜度数、散光度数及轴位。分别采用OPD-ScanⅢ和Pentacam行角膜生物测量且连续测量3次,记录平坦轴角膜曲率(K1)、陡峭轴角膜曲率(K2)、散光度数、散光轴位、散光矢量参数J0(0°或180°轴位的Jackson交叉柱镜值)、J45(45°轴位的Jackson交叉柱镜值)。采用组内相关系数(ICC)评价2种仪器测量数据的可重复性。分别采用Wilcoxon符号秩检验、Spearman秩相关性分析和Bland-Altman图表法分析2种仪器测量数据的差异、相关性及一致性。

结果OPD-Scan Ⅲ测得K1、K2、散光度数及轴位、J0以及J45的ICC均>0.900;Pentacam测得散光度数的ICC为0.896,其余测量数据均ICC>0.900。OPD-Scan Ⅲ测量K2、散光度数、J0及J45值明显高于Pentacam测量值,差异均有统计学意义(均 P<0.05)。OPD-Scan Ⅲ测量K1、K2、散光度数及轴位、J0和J45值与Pentacam的测量值均呈正相关( r s =0.981、0.982、0.900、0.737、0.921、0.703,均 P<0.01)。OPD-Scan Ⅲ与Pentacam测量K1、K2、散光度数及轴位、J0及J45差值的95%一致性界限(LOA)分别为-0.52~0.50 D、-0.39~0.59 D、-0.37~0.48 D、-17.29°~20.38°、-0.12~0.24 D和-0.22~0.28 D。

结论OPD-Scan Ⅲ在测量高度角膜散光患者的角膜屈光力及角膜散光度数时具有较好的可重复性和准确性,其测量散光轴位的95% LOA较大,不能被临床所接受。

角膜屈光力;角膜散光;OPD-Scan Ⅲ;可重复性;一致性
ABSTRACT

ObjectiveTo analyze the repeatability of keratometry and astigmatism values measured by the OPD-Scan Ⅲ and the agreement of the parameters measured by OPD-Scan Ⅲ and Pentacam.

MethodsA diagnostic test study design was adopted.Fifty patients (100 eyes) with refractive errors, aged from 21 to 35 years old, were selected from Second Affiliated Hospital of Nantong University and Lixiang Eye Hospital of Soochow University during August 2018.Spherical equivalent, astigmatim degree and axis were measured by Autorefraction.Corneal biometric measurements were measured three times continuously with the above two instruments.Keratometry values at the flat axis (K1), keratometry values at the steep axis (K2), astigmatim degree, axis, vector parameters J0 (Jackson cross cylinder at 0°or 180°) and J45 (Jackson cross cylinder at 45°) were recorded.Intraclass correlation coefficient (ICC) was used for repeatability analysis.Wilcoxon signed rank test, Spearman correlation analysis and Bland-Altman graphs were employed to analyze the comparability.This study adhered to the Declaration of Helsinki and was approved by an Ethics Committee of Lixiang Eye Hospital of Soochow University (No.SLER2018112). Written informed consent was obtained from each patient prior to any examination.

ResultsThe ICC of K1, K2, astigmatism, astigmatic axis, J0 and J45 measured by OPD-Scan Ⅲ were all greater than 0.900; the ICC of the astigmatism measured by Pentacam was 0.896, and the ICC of the other parameters measured by Pentacam were greater than 0.900; The values of K2, astigmatism, J0 and J45 measured by OPD-Scan Ⅲ were greater than those measured by Pentacam, and the differences were statistically significant (all at P<0.05). The values of K1, K2, astigmatism degree, axis, J0 and J45 measured by OPD-Scan Ⅲ were positively correlated with those measured by Pentacam (r s=0.981, 0.982, 0.900, 0.737, 0.921, 0.703, all at P<0.01). The 95% agreement of limits (LOA) of K1, K2, astigmatism, axis, J0 and J45 measurement difference between OPD-Scan Ⅲ and Pentacam were -0.52-0.50 D, -0.39-0.59 D, -0.37-0.48 D, -17.29°-20.38°, -0.12-0.24 D and -0.22-0.28 D, respectively.

ConclusionsOPD-Scan Ⅲ has high credibility in measuring corneal refractive power and astigmatism degree, but its 95% LOA of astigmatism axis is too large to be accepted clinically.

Keratometry;Corneal astigmatism;OPD-Scan Ⅲ;Reproducibility;Comparability
Song Yu, Email: mocdef.6ab21eyeluygnos;
引用本文

李盼盼,袁幽,黄黎黎,等. OPD-Scan Ⅲ测量角膜屈光力和散光的可重复性及其与Pentacam检测结果的一致性[J]. 中华实验眼科杂志,2021,39(03):238-242.

DOI:10.3760/cma.j.cn115989-20190323-00134

PERMISSIONS

Request permissions for this article from CCC.

评价本文
*以上评分为匿名评价
角膜占眼总屈光力的3/4,准确测量角膜屈光力、散光度数及轴位对角膜屈光手术参数设置和人工晶状体屈光度的计算(特别是散光型人工晶状体的型号及轴位的计算)等均具有重要意义 [ 1 , 2 ]。目前临床上测量角膜屈光力的常用仪器包括Pentacam三维眼前节分析仪、角膜曲率计、Lensar光学生物测量仪、VERION数字导航系统等。Pentacam在临床角膜屈光力测量上应用较为广泛,且已有大量研究证实Pentacam测量角膜屈光力及散光具有较高的准确性 [ 3 , 4 , 5 ]。OPD-ScanⅢ角膜/屈光分析仪是OPD-Scan系统的最新设备,集角膜地形图、波前像差、自动验光、眼内散光等多功能于一身 [ 6 , 7 , 8 , 9 ]。该仪器不但可以为临床医生提供更全面的信息,同时可以减少患者的检查时间,提高诊疗效率。但OPD-Scan Ⅲ在临床上应用时间较短,目前针对其测量结果准确性的研究较少。本研究主要对OPD-Scan Ⅲ测量角膜屈光力和散光的可重复性和准确性进行分析,为其临床应用提供依据。
参考文献
[1]
Jin H , Holzer MP , Rabsilber T et al. Intraocular lens power calculation after laser refractive surgery:corrective algorithm for corneal power estimation[J]J Cataract Refract Surg 201036(1):87-96. DOI: 10.1016/j.jcrs.2009.07.011 .
返回引文位置Google Scholar
百度学术
万方数据
[2]
Olsen T Calculation of intraocular lens power:a review[J]Acta Ophthalmol Scand 200785(5):472-485. DOI: 10.1111/j.1600-0420.2007.00879.x .
返回引文位置Google Scholar
百度学术
万方数据
[3]
Brunner M , Czanner G , Vinciguerra R et al. Improving precision for detecting change in the shape of the cornea in patients with keratoconus[J/OL]Sci Rep 20188(1):12345[2019-02-12]https://pubmed.ncbi.nlm.nih.gov/30120293/. DOI: 10.1038/s41598-018-30173-7 .
返回引文位置Google Scholar
百度学术
万方数据
[4]
Zhang H , Zhou H , Zhu T et al. Evaluation of Scheimpflug imaging parameters in blepharospasm and normal eyes[J/OL]BMC Ophthalmol 201818(1):23[2019-02-12]9. https://pubmed.ncbi.nlm.nih.gov/30185154/. DOI: 10.1186/s12886-018-0897-9 .
返回引文位置Google Scholar
百度学术
万方数据
[5]
Milla M , Piñero DP , Amparo F et al. Pachymetric measurements with a new Scheimpflug photography-based system:intraobserver repeatability and agreement with optical coherence tomography pachymetry[J]J Cataract Refract Surg 201137(2):310-316. DOI: 10.1016/j.jcrs.2010.08.038 .
返回引文位置Google Scholar
百度学术
万方数据
[6]
McGinnigle S , Naroo SA , Eperjesi F Evaluation of the auto-refraction function of the Nidek OPD-Scan Ⅲ[J]Clin Exp Optom 201497(2):160-163. DOI: 10.1111/cxo.12109 .
返回引文位置Google Scholar
百度学术
万方数据
[7]
Hamer CA , Buckhurst H , Purslow C et al. Comparison of reliability and repeatability of corneal curvature assessment with six keratometers[J]Clin Exp Optom 201699(6):583-589. DOI: 10.1111/cxo.12329 .
返回引文位置Google Scholar
百度学术
万方数据
[8]
Asgari S , Hashemi H OPD scan Ⅲ accuracy:Topographic and aberrometric indices after accelerated corneal cross-linking[J]J Curr Ophthalmol 201830(1):58-62. DOI: 10.1016/j.joco.2017.09.004 .
返回引文位置Google Scholar
百度学术
万方数据
[9]
He W , Qiu X , Zhang S et al. Comparison of long-term decentration and tilt in two types of multifocal intraocular lenses with OPD-Scan Ⅲ aberrometer[J]Eye (Lond) 201832(7):1237-1243. DOI: 10.1038/s41433-018-0068-5 .
返回引文位置Google Scholar
百度学术
万方数据
[10]
Naeser K Assessment and statistics of surgically induced astigmatism[J]Acta Ophthalmol 200886Suppl 1:5-28. DOI: 10.1111/j.1755-3768.2008.01234.x .
返回引文位置Google Scholar
百度学术
万方数据
[11]
Tan Q , Ng AL , Cheng GP et al. Repeatability of pupil size measurements with NIDEK OPD-Scan Ⅲ in myopic children[J/OL]Ophthalmic Physiol Opt 2020(2020-12-08)[2021-01-30]https://onlinelibrary.wiley.com/doi/full/10.1111/opo.12774. DOI: 10.1111/opo.12774 .
返回引文位置Google Scholar
百度学术
万方数据
[12]
Guilbert E , Saad A , Elluard M et al. Repeatability of keratometry measurements obtained with three topographers in keratoconic and normal corneas[J]J Refract Surg 201632(3):187-192. DOI: 10.3928/1081597X-20160113-01 .
返回引文位置Google Scholar
百度学术
万方数据
[13]
Asgari S , Hashemi H , Jafarzadehpur E et al. OPD-Scan Ⅲ:a repeatability and inter-device agreement study of a multifunctional device in emmetropia,ametropia,and keratoconus[J]Int Ophthalmol 201636(5):697-705. DOI: 10.1007/s10792-016-0189-4 .
返回引文位置Google Scholar
百度学术
万方数据
[14]
Eibschitz-Tsimhoni M , Tsimhoni O , Archer SM et al. Effect of axial length and keratometry measurement error on intraocular lens implant power prediction formulas in pediatric patients[J]J AAPOS 200812(2):173-176. DOI: 10.1016/j.jaapos.2007.10.012 .
返回引文位置Google Scholar
百度学术
万方数据
[15]
Viestenz A , Seitz B , Langenbucher A Evaluating the eye's rotational stability during standard photography:effect on determining the axial orientation of toric intraocular lenses[J]J Cataract Refract Surg 200531(3):557-561. DOI: 10.1016/j.jcrs.2004.07.019 .
返回引文位置Google Scholar
百度学术
万方数据
备注信息
A
宋愈,Email: mocdef.6ab21eyeluygnos
B
所有作者均声明不存在任何利益冲突
C
南通市卫生健康委员会课题项目 (MB2019005)
评论 (0条)
注册
登录
时间排序
暂无评论,发表第一条评论抢沙发
MedAI助手(体验版)
文档即答
智问智答
机器翻译
回答内容由人工智能生成,我社无法保证其准确性和完整性,该生成内容不代表我们的态度或观点,仅供参考。
生成快照
文献快照

你好,我可以帮助您更好的了解本文,请向我提问您关注的问题。

0/2000

《中华医学会杂志社用户协议》 | 《隐私政策》

《SparkDesk 用户协议》 | 《SparkDesk 隐私政策》

网信算备340104764864601230055号 | 网信算备340104726288401230013号

技术支持:

历史对话
本文全部
还没有聊天记录
设置
模式
纯净模式沉浸模式
字号