论著
ENGLISH ABSTRACT
非接触式睡眠呼吸监测仪在成人阻塞性睡眠呼吸暂停中的应用价值
薛健博
赵瑞
李静
赵龙
周兵
董霄松
韩芳
作者及单位信息
·
DOI: 10.3760/cma.j.cn112147-20210131-00088
Validation of a contact-free sleep apnea monitor in adults with obstructive sleep apnea
Xue Jianbo
Zhao Rui
Li Jing
Zhao Long
Zhou Bing
Dong Xiaosong
Han Fang
Authors Info & Affiliations
Xue Jianbo
Department of Respiratory and Critical Care Medicine, Peking University People′s Hospital, Beijing 100044, China
Zhao Rui
Department of Respiratory and Critical Care Medicine, Peking University People′s Hospital, Beijing 100044, China
Li Jing
Department of Respiratory and Critical Care Medicine, Peking University People′s Hospital, Beijing 100044, China
Zhao Long
Department of Respiratory and Critical Care Medicine, Peking University People′s Hospital, Beijing 100044, China
Zhou Bing
Department of Respiratory and Critical Care Medicine, Peking University People′s Hospital, Beijing 100044, China
Dong Xiaosong
Department of Respiratory and Critical Care Medicine, Peking University People′s Hospital, Beijing 100044, China
Han Fang
Department of Respiratory and Critical Care Medicine, Peking University People′s Hospital, Beijing 100044, China
·
DOI: 10.3760/cma.j.cn112147-20210131-00088
1321
183
0
1
2
2
PDF下载
APP内阅读
摘要

目的评价非接触式睡眠呼吸监测仪在成人阻塞性睡眠呼吸暂停(OSA)患者中的应用价值。

方法本研究选取2018年7月至2019年5月因打鼾就诊于北京大学人民医院睡眠中心的成人患者,在进行多导睡眠监测(PSG)的同时使用非接触式睡眠呼吸监测仪进行监测,利用均数检验、Pearson相关系数及Bland-Altman法评估非接触式睡眠呼吸监测仪生成的呼吸事件指数(REI)与PSG生成的呼吸暂停低通气指数(AHI)之间的一致性,计算非接触式睡眠呼吸监测仪诊断OSA的敏感度和特异度,并以受试者工作特征(ROC)曲线确定其诊断OSA的最佳界值。

结果共纳入198例患者,其中男120例,女78例,年龄45.5(33.0,57.3)岁。PSG生成的AHI与非接触式睡眠呼吸监测仪的REI差异无统计学意义[15.9(4.7,40.2)次/h比16.2(6.1,40.0)次/h, P=0.381),二者显著相关( r=0.914,P<0.05),其一致性检验的平均差-0.06次/h(95% CI:-18.44~18.31);如以REI≥5次/h作为诊断标准,则非接触式睡眠呼吸监测仪诊断OSA的敏感度为91.2%,特异度为58.0%;以REI≥13.3次/h诊断中重度OSA的敏感度、特异度分别为90.1%,71.1%。

结论该非接触式睡眠呼吸监测仪所得REI与PSG-AHI一致性较好,可用于OSA及中重度OSA患者的筛查。

睡眠呼吸暂停,阻塞性;多导睡眠监测;非接触式睡眠呼吸监测仪
ABSTRACT

ObjectiveTo evaluate the diagnostic value of a contact-free sleep apnea monitor in Chinese adults with obstructive sleep apnea (OSA).

MethodsOne hundred and ninety-eight participants with snoring were recruited between July 2018 and May 2019 in Sleep Center of Peking University People′s Hospital, using nocturnal polysomnography (PSG) and contact-free sleep apnea monitor simultaneously. We evaluated the difference between respiratory event index (REI) generated by contact-free sleep apnea monitor and PSG-Apnea-Hypopnea Index (AHI). We calculated the sensitivity and specificity of OSA diagnosis using the contact-free sleep apnea monitor by hypothesis testing for means, Pearson correlation coefficient and Bland-Altman plots. Then, we used the receiver operating characteristic (ROC) curve to find out the best cut-off of OSA diagnosis.

ResultsPSG-AHI and the REI of contact-free sleep apnea monitor showed no statistically significant difference [15.9 (4.7, 40.2) vs. 16.2 (6.1, 40.0), P=0.381], and they were significantly correlated ( r=0.914, P<0.05), with mean difference of -0.06 (95% CI:-18.44, 18.31). The ROC curve showed that if REI ≥5 events/h was used as diagnostic criteria, the sensitivity and specificity of diagnosing OSA were 91.2% and 58.0%, respectively. The sensitivity and specificity of the contact-free sleep apnea monitor REI≥13.3 in diagnosing moderate and severe OSA were 90.1% and 71.1%, respectively.

ConclusionThe REI obtained from the contact-free sleep apnea monitor showed a good agreement with the PSG-AHI, and therefore, the contact-free sleep apnea monitor can be used for the screening of patients with moderate and severe OSA.

Sleep apnea, obstructive;Polysomnography;Contact-free sleep apnea monitor
Dong Xiaosong, Email: nc.defudabe.hpukpgnosoaixgnod
引用本文

薛健博,赵瑞,李静,等. 非接触式睡眠呼吸监测仪在成人阻塞性睡眠呼吸暂停中的应用价值[J]. 中华结核和呼吸杂志,2021,44(10):880-885.

DOI:10.3760/cma.j.cn112147-20210131-00088

PERMISSIONS

Request permissions for this article from CCC.

评价本文
*以上评分为匿名评价
阻塞性睡眠呼吸暂停(obstructive sleep apnea,OSA)是一组因上气道塌陷导致睡眠过程中反复出现呼吸暂停或者低通气并因此引起一系列病理生理改变的临床症候群。OSA不仅可以引起夜间打鼾及白天嗜睡等症状,还可以增加心血管疾病 1 , 2 , 3、恶性肿瘤 4甚至猝死 5的风险。目前认为,诊断OSA的金标准仍是在专业睡眠中心进行的整夜多导睡眠监测(polysomnography,PSG),但该方法对设备及人员要求较高且费用昂贵,难以大规模开展 6;睡眠环境的改变和安装多种传感器也会影响患者睡眠,难以完全还原患者平日的睡眠情况。便携式睡眠监测(portable monitoring,PM)技术已被推荐为诊断中重度OSA的可替代监测技术 7,但也需要多个导联连接,无法完全排除传感器对患者睡眠情况的影响以及导联脱落信号丢失导致的诊断不准确。
近年来,由于电子工程学技术的发展,许多新型传感器和算法被应用于睡眠监测领域,其中不乏非接触式的监测系统。研究表明,安装有压力传感器的床垫对于OSA具有较高的诊断价值,并且受试者睡眠舒适度也有较大的改善 8 , 9 , 10;利用声压传感器对受试者呼吸运动中不同部位的声音及压力情况进行监测的Sonomat床垫在OSA诊断方面也表现出较高的可靠性 11。智能手机除可以作为数据存储与分析系统联结外部传感器 12 , 13外,还可以利用自身携带的扩音器、麦克风及陀螺仪收集周围环境声音及体动信号进行分析以实现对睡眠状态的监测及疾病的诊断 14 , 15。利用低功率电磁波的雷达系统对睡眠呼吸疾病进行监测及诊断的SleepMinder也应用于临床,研究表明其在OSA诊断方面有着较高的准确性 16 , 17
本次研究中应用的非接触式睡眠呼吸监测仪是基于超宽带(impulse radio-ultra wide band,IR-UWB)雷达系统应用于OSA诊断的一种全新技术,其传感器系统与患者完全分离,减少了监测对患者睡眠情况的影响,但其准确性及诊断价值尚需进一步验证。本研究将非接触式睡眠呼吸监测仪与PSG进行对比,探讨其在成人OSA患者中的应用价值。
试读结束,您可以通过登录机构账户或个人账户后获取全文阅读权限。
参考文献
[1]
Javaheri S , Barbe F , Campos-Rodriguez F ,et al. Sleep apnea: types, mechanisms, and clinical cardiovascular consequences[J]. J Am Coll Cardiol, 2017,69(7):841-858. DOI: 10.1016/j.jacc.2016.11.069 .
返回引文位置Google Scholar
百度学术
万方数据
[2]
Eastwood PR , Malhotra A , Palmer LJ ,et al. Obstructive Sleep Apnoea: From pathogenesis to treatment: Current controversies and future directions[J]. Respirology, 2010,15(4):587-595. DOI: 10.1111/j.1440-1843.2009.01699.x .
返回引文位置Google Scholar
百度学术
万方数据
[3]
Park JG , Ramar K , Olson EJ . Updates on definition, consequences, and management of obstructive sleep apnea[J]. Mayo Clin Proc, 2011,86(6):549-554; quiz 554-555. DOI: 10.4065/mcp.2010.0810 .
返回引文位置Google Scholar
百度学术
万方数据
[4]
Campos-Rodriguez F , Martinez-Garcia MA , Martinez M ,et al. Association between obstructive sleep apnea and cancer incidence in a large multicenter Spanish cohort[J]. Am J Respir Crit Care Med, 2013,187(1):99-105. DOI: 10.1164/rccm.201209-1671OC .
返回引文位置Google Scholar
百度学术
万方数据
[5]
Gami AS , Olson EJ , Shen WK ,et al. Obstructive sleep apnea and the risk of sudden cardiac death: a longitudinal study of 10, 701 adults[J]. J Am Coll Cardiol, 2013,62(7):610-616. DOI: 10.1016/j.jacc.2013.04.080 .
返回引文位置Google Scholar
百度学术
万方数据
[6]
卢迷,房芳,魏永祥. 便携式监测设备在成人阻塞性睡眠呼吸暂停低通气综合征诊断中的临床应用[J]. 中华耳鼻咽喉头颈外科杂志, 2019,54(6):477-480. DOI: 10.3760/cma.j.issn.1673-0860.2019.06.016 .
返回引文位置Google Scholar
百度学术
万方数据
[7]
Kapur VK , Auckley DH , Chowdhuri S ,et al. Clinical practice guideline for diagnostic testing for adult obstructive sleep apnea: an American Academy of Sleep Medicine clinical practice guideline[J]. J Clin Sleep Med, 2017,13(3):479-504. DOI: 10.5664/jcsm.6506 .
返回引文位置Google Scholar
百度学术
万方数据
[8]
Kobayashi M , Namba K , Tsuiki S ,et al. Validity of sheet-type portable monitoring device for screening obstructive sleep apnea syndrome[J]. Sleep Breath, 2013,17(2):589-595. DOI: 10.1007/s11325-012-0725-z .
返回引文位置Google Scholar
百度学术
万方数据
[9]
Takasaki Y , Kaneko Y , Sakakibara H ,et al. Clinical usefulness and health-economic benefits of a new sheet-like medical device (SD-101) for the diagnosis of sleep apnea syndrome[J]. Nihon Kokyuki Gakkai Zasshi, 2008,46(3):181-188.
返回引文位置Google Scholar
百度学术
万方数据
[10]
Meng L , Xu H , Guan J ,et al. Validation of a novel sleep-monitoring system for diagnosing obstructive sleep apnea: a comparison with polysomnography[J]. Exp Ther Med, 2016,12(5):2937-2941. DOI: 10.3892/etm.2016.3721 .
返回引文位置Google Scholar
百度学术
万方数据
[11]
Norman MB , Middleton S , Erskine O ,et al. Validation of the Sonomat: a contactless monitoring system used for the diagnosis of sleep disordered breathing[J]. Sleep, 2014,37(9):1477-1487. DOI: 10.5665/sleep.3996 .
返回引文位置Google Scholar
百度学术
万方数据
[12]
Zhang J , Zhang Q , Wang Y ,et al. A real-time auto-adjustable smart pillow system for sleep apnea detection and treatment[C]//2013 ACM/IEEE International Conference on Information Processing in Sensor Networks (IPSN),IEEE, 2013:179-190.
返回引文位置Google Scholar
百度学术
万方数据
[13]
Garde A , Karlen W , Dehkordi P ,et al. Oxygen saturation in children with and without obstructive sleep apnea using the phone-oximeter[J]. Annu Int Conf IEEE Eng Med Biol Soc, 2013,2013:2531-2534. DOI: 10.1109/EMBC.2013.6610055 .
返回引文位置Google Scholar
百度学术
万方数据
[14]
Nandakumar R , Gollakota S , Watson N . Contactless sleep apnea detection on smartphones[C]//Proceedings of the 13th annual international conference on mobile systems,applications, and services, 2015:45-57.
返回引文位置Google Scholar
百度学术
万方数据
[15]
Zhao Y , Zhang H , Liu W ,et al. A snoring detector for OSAHS based on patient′s individual personality[C]//2011 3rd International Conference on Awareness Science and Technology (iCAST),IEEE, 2011:24-27.
返回引文位置Google Scholar
百度学术
万方数据
[16]
De Chazal P , Fox N , O′Hare E ,et al. Sleep/wake measurement using a non-contact biomotion sensor[J]. J Sleep Res, 2011,20(2):356-366. DOI: 10.1111/j.1365-2869.2010.00876.x .
返回引文位置Google Scholar
百度学术
万方数据
[17]
Crinion SJ , Tiron R , Lyon G ,et al. Ambulatory detection of sleep apnea using a non-contact biomotion sensor[J]. J Sleep Res, 2020,29(1):e12889. DOI: 10.1111/jsr.12889 .
返回引文位置Google Scholar
百度学术
万方数据
[18]
中华医学会呼吸病学分会睡眠呼吸障碍学组. 阻塞性睡眠呼吸暂停低通气综合征诊治指南(2011年修订版)[J]. 中华结核和呼吸杂志, 2012,35(1):9-12. DOI: 10.3760/cma.j.issn.1001-0939.2012.01.007 .
返回引文位置Google Scholar
百度学术
万方数据
[19]
Berry RB , Brooks R , Gamaldo CE ,et al. The AASM manual for the scoring of sleep and associated events[J]. Rules, Terminology and Technical Specifications, Version 2.6. Darien, Illinois, American Academy of Sleep Medicine, 2020.
返回引文位置Google Scholar
百度学术
万方数据
[20]
Lazaro A , Girbau D , Villarino R . Analysis of vital signs monitoring using an IR-UWB radar[J]. Progress In Electromagnetics Research, 2010,100:265-284.
返回引文位置Google Scholar
百度学术
万方数据
[21]
陈卉. Bland-Altman分析在临床测量方法一致性评价中的应用[J]. 中国卫生统计, 2007,24(3):308-309, 315. DOI: 10.3969/j.issn.1002-3674.2007.03.029 .
返回引文位置Google Scholar
百度学术
万方数据
[22]
Pallesen S , Grønli J , Myhre K ,et al. A pilot study of impulse radio ultra wideband radar technology as a new tool for sleep assessment[J]. J Clin Sleep Med, 2018,14(7):1249-1254. DOI: 10.5664/jcsm.7236 .
返回引文位置Google Scholar
百度学术
万方数据
[23]
Wang D , Yoo S , Cho SH . Experimental comparison of IR-UWB radar and FMCW radar for vital signs[J]. Sensors (Basel), 2020,20(22). DOI: 10.3390/s20226695 .
返回引文位置Google Scholar
百度学术
万方数据
[24]
Tantiparimongkol L , Phasukkit P . IR-UWB pulse generation using FPGA scheme for through obstacle human detection[J]. Sensors (Basel), 2020,20(13). DOI: 10.3390/s20133750 .
返回引文位置Google Scholar
百度学术
万方数据
[25]
Bushberg JT , Tupin JP . RF Safety analysis of a novel ultra-wideband fetal monitoring system[J]. Health Phys, 2017,112(5):478-485. DOI: 10.1097/HP.0000000000000649 .
返回引文位置Google Scholar
百度学术
万方数据
[26]
Federal communications commission, electronic code of federal regulations: Title 47, Part 15, Subpart F, American National Standards Institute C63.10:2013.
[27]
Membership of the Advisory Group on Non-ionising Radiation. Health effects from radiofrequency electromagnetic fields. Documents of the Health Protection Agency Radiation, Chemical and Environmental Hazards. 2012(978-0-85951-714-0).
返回引文位置Google Scholar
百度学术
万方数据
[28]
冯晨,张惠栋,韩莹莹,. 脉冲无线电超宽带雷达技术在成人OSA中的应用研究[J]. 临床耳鼻咽喉头颈外科杂志, 2020,34(7):634-638. DOI: 10.13201/j.issn.2096-7993.2020.07.014 .
返回引文位置Google Scholar
百度学术
万方数据
[29]
Zaffaroni A , Kent B , O′Hare E ,et al. Assessment of sleep-disordered breathing using a non-contact bio-motion sensor[J]. J Sleep Res, 2013,22(2):231-236. DOI: 10.1111/j.1365-2869.2012.01056.x .
返回引文位置Google Scholar
百度学术
万方数据
[30]
Xu L , Han F , Keenan BT ,et al. Validation of the Nox-T3 portable monitor for diagnosis of obstructive sleep apnea in Chinese adults[J]. J Clin Sleep Med, 2017,13(5):675-683. DOI: 10.5664/jcsm.6582 .
返回引文位置Google Scholar
百度学术
万方数据
[31]
Morales CR , Hurley S , Wick LC ,et al. In-home, self-assembled sleep studies are useful in diagnosing sleep apnea in the elderly[J]. Sleep, 2012,35(11):1491-1501. DOI: 10.5665/sleep.2196 .
返回引文位置Google Scholar
百度学术
万方数据
[32]
Chai-Coetzer CL , Antic NA , Rowland LS ,et al. A simplified model of screening questionnaire and home monitoring for obstructive sleep apnoea in primary care[J]. Thorax, 2011,66(3):213-219. DOI: 10.1136/thx.2010.152801 .
返回引文位置Google Scholar
百度学术
万方数据
[33]
Gjevre JA , Taylor-Gjevre RM , Skomro R ,et al. Comparison of polysomnographic and portable home monitoring assessments of obstructive sleep apnea in Saskatchewan women[J]. Can Respir J, 2011,18(5):271-274. DOI: 10.1155/2011/408091 .
返回引文位置Google Scholar
百度学术
万方数据
备注信息
A
董霄松,Email: nc.defudabe.hpukpgnosoaixgnod
B

薛健博、赵瑞:受试者入组、数据收集、数据分析、论文撰写;李静、赵龙、周兵:PSG数据采集和分析;董霄松、韩芳:研究设计、研究指导、论文修改

C

薛健博, 赵瑞, 李静, 等. 非接触式睡眠呼吸监测仪在成人阻塞性睡眠呼吸暂停中的应用价值[J]. 中华结核和呼吸杂志, 2021, 44(10): 880-885. DOI: 10.3760/cma.j.cn112147-20210131-00088.

D
所有作者均声明不存在利益冲突
E
国家高技术研究发展计划 (2012AA02A604)
评论 (0条)
注册
登录
时间排序
暂无评论,发表第一条评论抢沙发
MedAI助手(体验版)
文档即答
智问智答
机器翻译
回答内容由人工智能生成,我社无法保证其准确性和完整性,该生成内容不代表我们的态度或观点,仅供参考。
生成快照
文献快照

你好,我可以帮助您更好的了解本文,请向我提问您关注的问题。

0/2000

《中华医学会杂志社用户协议》 | 《隐私政策》

《SparkDesk 用户协议》 | 《SparkDesk 隐私政策》

网信算备340104764864601230055号 | 网信算备340104726288401230013号

技术支持:

历史对话
本文全部
还没有聊天记录
设置
模式
纯净模式沉浸模式
字号