临床研究
ENGLISH ABSTRACT
CASIA2、IOLMaster 700和Pentacam对白内障眼角膜屈光力及散光测量的比较
王婷婷
王抒阳
张敏
秦鲁宁
林松
作者及单位信息
·
DOI: 10.3760/cma.j.cn115989-20210730-00435
Comparison of corneal refractive power and astigmatism measured by CASIA2, IOLMaster 700 and Pentacam in cataract patients
Wang Tingting
Wang Shuyang
Zhang Min
Qin Luning
Lin Song
Authors Info & Affiliations
Wang Tingting
Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin Branch of National Clinical Research Center for Ocular Disease, Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital, Tianjin 300384, China
Wang Shuyang
Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin Branch of National Clinical Research Center for Ocular Disease, Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital, Tianjin 300384, China
Zhang Min
Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin Branch of National Clinical Research Center for Ocular Disease, Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital, Tianjin 300384, China
Qin Luning
Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin Branch of National Clinical Research Center for Ocular Disease, Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital, Tianjin 300384, China
Lin Song
Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin Branch of National Clinical Research Center for Ocular Disease, Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital, Tianjin 300384, China
·
DOI: 10.3760/cma.j.cn115989-20210730-00435
0
0
0
0
0
0
PDF下载
APP内阅读
摘要

目的比较CASIA2、IOLMaster 700和Pentacam测量年龄相关性白内障患者角膜前表面和全角膜屈光力及散光的差异性和一致性。

方法采用诊断试验研究方法,收集2021年3—4月在天津医科大学眼科医院就诊的年龄相关性白内障患者200例200眼,分别采用CASIA2、IOLMaster 700和Pentacam测量角膜前表面及全角膜的陡峭轴角膜曲率(Ks)、平坦轴角膜曲率(Kf)及平均角膜曲率(Km)、散光度数及轴向。采用傅里叶转换公式将散光矢量转换成J0和J45。3种仪器测量值的差异和相关性分析分别采用重复测量单因素方差分析和Pearson相关分析,一致性评估采用Bland-Altman一致性检验。

结果CASIA2、IOLMaster 700和Pentacam测量的角膜前表面Kf和J0总体比较差异均有统计学意义( F=18.563、16.172,均 P=0.001),CASIA2测量Kf值大于IOLMaster 700,IOLMaster 700测量J0值明显大于Pentacam,差异均有统计学意义(均 P<0.05)。各仪器测量全角膜Ks、Kf、Km、J0总体比较差异均有统计学意义( F=1 300.447、1 274.117、1 609.713、10.372,均 P=0.001),其中IOLMaster 700测量值最大,CASIA2次之,Pentacam最小,两两比较差异均有统计学意义(均 P<0.05)。Pearson相关分析显示,3种仪器测量角膜屈光力具有较强的相关性(均 r>0.935, P<0.01),散光度数值相关性均弱于角膜屈光力测量值( r=0.623~0.908,均 P<0.01)。Bland-Altman分析显示3种仪器测量角膜前表面屈光力、角膜前表面散光及全角膜散光有较好的一致性,均在临床可接受范围,而测量全角膜屈光力的一致性较差。其中,IOLMaster 700与Pentacam测量全角膜屈光力值差异较大,CASIA2与Pentacam差异相对较小。

结论CASIA2、IOLMaster 700和Pentacam测量年龄相关性白内障患者角膜前表面屈光力及散光一致性较好,可以互相参考,而测量全角膜屈光力的一致性较差,其中IOLMaster 700测量值最大,Pentacam测量值最小,临床上不建议互换。

白内障;体层摄影术,光学相干;角膜地形图;散光;角膜屈光力
ABSTRACT

ObjectiveTo compare the difference and consistency of anterior corneal surface and total corneal refractive power and astigmatism measured by CASIA2, IOLMaster 700 and Pentacam in patients with age-related cataract.

MethodsA diagnostic test was conducted.Two-hundred patients (200 eyes) with age-related cataract were enrolled in Tianjin Medical University Eye Hospital from March to April 2021.The steep keratometry (Ks), flat keratometry (Kf), mean keratometry (Km), degree and axis of astigmatism of the anterior and the total corneal surface of patients were measured by CASIA2, IOLMaster 700 and Pentacam, respectively.The astigmatism was transformed into J0 and J45 by Fourier transform formula.The differences and correlation of the measurements obtained with the three instruments were analyzed by one-way repeated measures analysis of variance and Pearson correlation analysis.The consistency was evaluated by Bland-Altman test.This study adhered to the Declaration of Helsinki.The study protocol was approved by the Ethics Committee of Tianjin Medical University Eye Hospital (No.2021KY-07).

ResultsThere were statistically significant differences in anterior corneal surface Kf and J0 measured by the three instruments ( F=18.563, 16.172; both at P=0.001). The Kf measured by CASIA2 was significantly higher than that measured by IOLMaster 700, and the J0 measured by IOLMaster 700 was significantly higher than that measured by Pentacam (both at P<0.05). There were statistically significant differences in total corneal Ks, Kf, Km and J0 measured by the three instruments, which from IOLMaster 700 were the largest, followed by CASIA2, then Pentacam ( F=1 300.447, 1 274.117, 1 609.713, 10.372; all at P=0.001). Pearson correlation analysis showed that the corneal refractive power measured by the three instruments was highly correlated (all at r>0.935, P<0.01), and the correlation of astigmatism values was weaker than the corneal refractive power ( r=0.623-0.908, all at P<0.01). Bland-Altman analysis showed that the three instruments had good consistency in measuring the anterior corneal surface refractive power, anterior corneal surface astigmatism and total corneal astigmatism, which were clinically acceptable, while the consistency of total corneal refractive power measurement was poor.The difference in measuring total corneal refractive power was large between IOLMaster 700 and Pentacam, and relatively small between CASIA2 and Pentacam.

ConclusionsThe consistency of CASIA2, IOLMaster 700 and Pentacam is good in measuring the anterior corneal surface refractive power of patients with age-related cataract, which can be substitutable, but poor in measuring the total corneal refractive power.The total corneal refractive power measurement from IOLMaster700 is the largest, the smallest from Pentacam, which is not recommended to be clinically exchangeable.

Cataract;Tomography, optical coherence;Corneal topography;Astigmatism;Corneal refractive power
Lin Song, Email: mocdef.aabnis321321gnosnil
引用本文

王婷婷,王抒阳,张敏,等. CASIA2、IOLMaster 700和Pentacam对白内障眼角膜屈光力及散光测量的比较[J]. 中华实验眼科杂志,2022,40(11):1055-1061.

DOI:10.3760/cma.j.cn115989-20210730-00435

PERMISSIONS

Request permissions for this article from CCC.

评价本文
*以上评分为匿名评价
随着角膜屈光手术和人工晶状体(intraocular lens,IOL)植入术的不断发展,精确的角膜生物学测量对术后视觉质量的提高至关重要。术前角膜屈光力和散光的精准测量是精确计算IOL屈光力的前提,也是术后获得良好视觉质量的基础。全角膜屈光力包括角膜前表面屈光力和角膜后表面屈光力。以往仅用角膜前表面屈光力和角膜屈光指数模拟全角膜屈光力。近年来,随着生物测量仪器的发展,可以真实地测量角膜前、后表面的屈光力及散光,为IOL屈光力的计算,特别是Toric IOL等其他高端IOL屈光力的计算及手术切口选择提供更准确的手术依据。三维眼前节分析仪Pentacam使用Scheimpflug相机旋转摄像获取三维眼前节图像,可提供不同直径范围环上和区域内的前后表面角膜屈光力、净屈光力、总角膜屈光力等数据,已广泛应用于临床Toric IOL屈光力计算。近年,新一代扫频源眼前节光相干断层扫描仪CASIA2问世并受到广泛关注,其应用低相干光干涉原理实现眼前节断层结构显示及生物学测量,具有更快的扫描速度和更高的分辨率,可提供角膜前后表面及全角膜的屈光力及散光、前房结构、晶状体厚度等方面的数据 [ 1 ]。IOLMaster 700是一种扫频光学生物测量仪,具有高信噪比、长波长、更好的组织穿透力及更快的扫描速度,能测量角膜屈光力、中央角膜厚度、前房深度、白到白距离、瞳孔直径、眼轴长度和晶状体厚度,被认为是IOL屈光力计算的"金标准" [ 2 ];其最新版本的测量软件增加了角膜后表面和全角膜屈光力及散光的测量。由于CASIA2设备及IOLMaster 700最新版软件均刚投入使用,其测量全角膜屈光力的准确性研究较少。本研究拟分析CASIA2、IOLMaster 700及Pentacam测量角膜前表面和全角膜屈光力及散光的差异性、相关性及一致性,为其临床应用提供参考。
试读结束,您可以通过登录机构账户或个人账户后获取全文阅读权限。
参考文献
[1]
Chansangpetch S Nguyen A Mora M et al. Agreement of anterior segment parameters obtained from swept-source Fourier-domain and time-domain anterior segment optical coherence tomography[J]Invest Ophthalmol Vis Sci 201859(3)∶15541561. DOI: 10.1167/iovs.17-23574 .
返回引文位置Google Scholar
百度学术
万方数据
[2]
Lee AC Qazi MA Pepose JS . Biometry and intraocular lens power calculation[J]Curr Opin Ophthalmol 200819(1)∶1317. DOI: 10.1097/ICU.0b013e3282f1c5ad .
返回引文位置Google Scholar
百度学术
万方数据
[3]
Michelitsch M Ardjomand N Vidic B et al. Prevalence and age-related changes of corneal astigmatism in patients before cataract surgery[J]Ophthalmologe 2017114(3)∶247251. DOI: 10.1007/s00347-016-0323-8 .
返回引文位置Google Scholar
百度学术
万方数据
[4]
Day AC Dhariwal M Keith MS et al. Distribution of preoperative and postoperative astigmatism in a large population of patients undergoing cataract surgery in the UK[J]Br J Ophthalmol 2019103(7)∶9931000. DOI: 10.1136/bjophthalmol-2018-312025 .
返回引文位置Google Scholar
百度学术
万方数据
[5]
Villegas EA Alcón E Artal P Minimum amount of astigmatism that should be corrected[J]J Cataract Refract Surg 201440(1)∶1319. DOI: 10.1016/j.jcrs.2013.09.010 .
返回引文位置Google Scholar
百度学术
万方数据
[6]
Koch DD Ali SF Weikert MP et al. Contribution of posterior corneal astigmatism to total corneal astigmatism[J]J Cataract Refract Surg 201238(12)∶20802087. DOI: 10.1016/j.jcrs.2012.08.036 .
返回引文位置Google Scholar
百度学术
万方数据
[7]
Tonn B Klaproth OK Kohnen T Anterior surface-based keratometry compared with Scheimpflug tomography-based total corneal astigmatism[J]Invest Ophthalmol Vis Sci 201456(1)∶291298. DOI: 10.1167/iovs.14-15659 .
返回引文位置Google Scholar
百度学术
万方数据
[8]
Ho JD Tsai CY Liou SW . Accuracy of corneal astigmatism estimation by neglecting the posterior corneal surface measurement[J]Am J Ophthalmol 2009147(5)∶788795. DOI: 10.1016/j.ajo.2008.12.020 .
返回引文位置Google Scholar
百度学术
万方数据
[9]
Davison JA Potvin R Refractive cylinder outcomes after calculating t oric intraocular lens cylinder power using total corneal refractive power [J]Clin Ophthalmol 2015915111517. DOI: 10.2147/OPTH.S88693 .
返回引文位置Google Scholar
百度学术
万方数据
[10]
Srivannaboon S Chirapapaisan C Comparison of refractive outcomes using conventional keratometry or total keratometry for IOL power calculation in cataract surgery[J] Graefe ' s Arch Clin Exp Ophthalmol 2019257(12)∶26772682. DOI: 10.1007/s00417-019-04443-7 .
返回引文位置Google Scholar
百度学术
万方数据
[11]
Piñero DP Soto-Negro R Ruiz-Fortes P et al. Analysis of intrasession repeatability of ocular aberrometric measurements and validation of keratometry provided by a new integrated system in mild to moderate keratoconus[J]Cornea 201938(9)∶10971104. DOI: 10.1097/ICO.0000000000002034 .
返回引文位置Google Scholar
百度学术
万方数据
[12]
Kawamorita T Nakayama N Uozato H Repeatability and reproducibility of corneal curvature measurements using the Pentacam and Keratron topography systems[J]J Refract Surg 200925(6)∶539544. DOI: 10.3928/1081597X-20090512-08 .
返回引文位置Google Scholar
百度学术
万方数据
[13]
Srivannaboon S Chirapapaisan C Chonpimai P et al. Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer[J]J Cataract Refract Surg 201541(10)∶22242232. DOI: 10.1016/j.jcrs.2015.03.019 .
返回引文位置Google Scholar
百度学术
万方数据
[14]
Sel S Stange J Kaiser D et al. Repeatability and agreement of Scheimpflug-based and swept-source optical biometry measurements[J]Cont Lens Anterior Eye 201740(5)∶318322. DOI: 10.1016/j.clae.2017.03.007 .
返回引文位置Google Scholar
百度学术
万方数据
[15]
Zhang T Zhou Y Young CA et al. Comparison of a new swept-source anterior segment optical coherence tomography and a Scheimpflug camera for measurement of corneal curvature[J]Cornea 202039(7)∶818822. DOI: 10.1097/ICO.0000000000002280 .
返回引文位置Google Scholar
百度学术
万方数据
[16]
Xu BY Mai DD Penteado RC et al. Reproducibility and agreement of anterior segment parameter measurements obtained using the CASIA2 and Spectralis OCT2 optical coherence tomography devices[J]J Glaucoma 201726(11)∶974979. DOI: 10.1097/IJG.0000000000000788 .
返回引文位置Google Scholar
百度学术
万方数据
[17]
Jin GM Xiao B Zhou YJ et al. Agreement of corneal curvature and central corneal thickness obtained from a swept-source OCT and Pentacam in ectopia lentis patients[J]Int J Ophthalmol 2020(13)∶81244. 1249DOI: 10.18240/ijo.2020.08.10 .
返回引文位置Google Scholar
百度学术
万方数据
[18]
高奕晨蒋元丰林松新型眼前节相干光层析成像仪与Scheimpflug眼前节分析仪测量年龄相关性白内障患者角膜屈光力及散光的比较[J]中华眼科杂志 202157(1)∶4855. DOI: 10.3760/cma.j.cn112142-20200904-00574 .
返回引文位置Google Scholar
百度学术
万方数据
Gao YC Jiang YF Lin S et al. Comparison of corneal refractive power and astigmatism measured by the new anterior segment optical coherence tomographic device and Scheimpflug imaging device in age-related cataract patients[J]Chin J Ophthalmol 202157(1)∶4855. DOI: 10.3760/cma.j.cn112142-20200904-00574 .
Goto CitationGoogle Scholar
Baidu Scholar
Wanfang Data
[19]
Szalai E Berta A Hassan Z et al. Reliability and repeatab ility of swept-source Fourier-domain optical coherence tomography and Scheimpflug imaging in keratoconus [J]J Cataract Refract Surg 201238(3)∶485494. DOI: 10.1016/j.jcrs.2011.10.027 .
返回引文位置Google Scholar
百度学术
万方数据
[20]
Özyol P Özyol E Agreement between swept-source optical biometry and Scheimpflug-based topography measurements of anterior segment parameters[J]Am J Ophthalmol 20161697378. DOI: 10.1016/j.ajo.2016.06.020 .
返回引文位置Google Scholar
百度学术
万方数据
[21]
Shajari M Sonntag R Ramsauer M et al. Evaluation of total corneal power measurements with a new optical biometer[J]J Cataract Refract Surg 202046(5)∶675681. DOI: 10.1097/j.jcrs.0000000000000136 .
返回引文位置Google Scholar
百度学术
万方数据
[22]
Dong J Tang M Zhang Y et al. Comparison of anterior segment biometric measurements between Pentacam HR and IOLMaster in normal and high myopic eyes[J/OL]PLoS One 201510(11)∶e0143110[2021-11-06]http://www.ncbi.nlm.nih.gov/pubmed/26575265. DOI: 10.1371/journal.pone.0143110 .
返回引文位置Google Scholar
百度学术
万方数据
[23]
Kiraly L Stange J Kunert KS et al. Repeatability and agreement of central corneal thickness and keratometry measurements between four different devices[J/OL]J Ophthalmol 201720176181405[2021-11-06]http://www.ncbi.nlm.nih.gov/pubmed/28357136. DOI: 10.1155/2017/6181405 .
返回引文位置Google Scholar
百度学术
万方数据
[24]
Molina-Martín A Piñero DP Caballero MT et al. Comparative analysis of anterior corneal curvature and astigmatism measurements obtained with three different devices[J]Clin Exp Optom 2020103(5)∶618624. DOI: 10.1111/cxo.13002 .
返回引文位置Google Scholar
百度学术
万方数据
[25]
Montalbán R Piñero DP Javaloy J et al. Scheimpflug photography-based clinical characterization of the correlation of the corneal shape between the anterior and posterior corneal surfaces in the normal human eye[J]J Cataract Refract Surg 201238(11)∶19251933. DOI: 10.1016/j.jcrs.2012.06.050 .
返回引文位置Google Scholar
百度学术
万方数据
[26]
Fesharaki H Azizzadeh A Ghoreishi SM et al. The effects of lateral head tilt on ocular astigmatic axis[J/OL]Adv Biomed Res 2014310[2021-11-28]http://www.ncbi.nlm.nih.gov/pubmed/24592363. DOI: 10.4103/2277-9175.124638 .
返回引文位置Google Scholar
百度学术
万方数据
[27]
洪莲花李英俊干眼对年龄相关性白内障患者人工晶状体度数测量和计算准确性的影响[J]中华实验眼科杂志 201634(2)∶170174. DOI: 10.3760/cma.j.issn.2095-0160.2016.02.016 .
返回引文位置Google Scholar
百度学术
万方数据
Hong LH Li YJ . Pilot study on dry eye affecting the accuracy of intraocular lens power measurement in age-related cataract[J]Chin J Exp Ophthalmol 201634(2)∶170174. DOI: 10.3760/cma.j.issn.2095-0160.2016.02.016 .
Goto CitationGoogle Scholar
Baidu Scholar
Wanfang Data
[28]
Karunaratne N Comparison of the Pentacam equivalent keratometry reading and IOL Master keratometry measurement in intraocular lens power calculations[J]Clin Exp Ophthalmol 201341(9)∶825834. DOI: 10.1111/ceo.12124 .
返回引文位置Google Scholar
百度学术
万方数据
备注信息
A
林松,Email: mocdef.aabnis321321gnosnil
B

王婷婷:酝酿和设计试验、实施研究、采集数据、分析/解释数据,文章撰写;王抒阳、张敏、秦鲁宁:实施研究、采集数据、分析/解释数据;林松:酝酿和设计试验、指导试验、对文章的知识性内容作批评性审阅

C
所有作者均声明不存在利益冲突
评论 (0条)
注册
登录
时间排序
暂无评论,发表第一条评论抢沙发
MedAI助手(体验版)
文档即答
智问智答
机器翻译
回答内容由人工智能生成,我社无法保证其准确性和完整性,该生成内容不代表我们的态度或观点,仅供参考。
生成快照
文献快照

你好,我可以帮助您更好的了解本文,请向我提问您关注的问题。

0/2000

《中华医学会杂志社用户协议》 | 《隐私政策》

《SparkDesk 用户协议》 | 《SparkDesk 隐私政策》

网信算备340104764864601230055号 | 网信算备340104726288401230013号

技术支持:

历史对话
本文全部
还没有聊天记录
设置
模式
纯净模式沉浸模式
字号