临床研究
ENGLISH ABSTRACT
基于扫频源OCT技术光学生物测量仪对IOL倾斜测量的可行性评价及倾斜特征分析
赵琦
杨文利
李栋军
王子杨
陈伟
李逸丰
崔蕊
沈琳
李蕾
鲜军舫
作者及单位信息
·
DOI: 10.3760/cma.j.cn115989-20210311-00167
Analysis of the feasibility of measurements and characteristics of intraocular lens tilt using swept-source optical coherence tomography biometer
Zhao Qi
Yang Wenli
Li Dongjun
Wang Ziyang
Chen Wei
Li Yifeng
Cui Rui
Shen Lin
Li Lei
Xian Junfang
Authors Info & Affiliations
Zhao Qi
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Science Key Laboratory, Beijing 100730, China
Yang Wenli
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Science Key Laboratory, Beijing 100730, China
Li Dongjun
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Science Key Laboratory, Beijing 100730, China
Wang Ziyang
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Science Key Laboratory, Beijing 100730, China
Chen Wei
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Science Key Laboratory, Beijing 100730, China
Li Yifeng
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Science Key Laboratory, Beijing 100730, China
Cui Rui
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Science Key Laboratory, Beijing 100730, China
Shen Lin
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Science Key Laboratory, Beijing 100730, China
Li Lei
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Science Key Laboratory, Beijing 100730, China
Xian Junfang
Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China
·
DOI: 10.3760/cma.j.cn115989-20210311-00167
0
0
0
0
0
0
PDF下载
APP内阅读
摘要

目的分析基于扫频源光学相干断层扫描(OCT)技术光学生物测量仪IOLMaster 700评估人工晶状体(IOL)倾斜测量的可行性、IOL倾斜特征及相关影响因素。

方法采用横断面研究,纳入2018年7—9月于首都医科大学附属北京同仁医院行白内障超声乳化吸除联合IOL植入术术后患者185例242眼,其中右眼119眼,左眼123眼。采用IOLMaster 700扫描获取眼内IOL位置图像及κ角、α角、角膜曲率、前房深度(ACD)、瞳孔直径(PD)、眼轴长度(AL)等,计算IOL倾斜方向及幅度。采用合并组内相关标准差及组内相关系数计算法评估3次IOL倾斜测量的可重复性。采用Pearson线性相关分析或Spearman秩相关分析评估IOL倾斜的双眼对称性及IOL倾斜与不同影响因素之间的相关关系。IOL倾斜影响因素评估采用多元线性回归分析法。

结果IOL倾斜测量具有较好的可重复性,合并组内相关标准差IOL倾斜方向为5.16°,IOL倾斜幅度为0.13°;双眼IOL倾斜具有镜像对称性,右眼IOL倾斜方向为-89.60~87.53°,平均(18.85±29.65)°,倾斜幅度为1.32~7.05°,平均(4.16±1.26)°。左眼IOL倾斜方向为-84.30~89.44°,平均(21.17±32.38)°,倾斜幅度为0.58~7.40°,平均(3.80±1.31)°。右眼和左眼IOL倾斜方向和倾斜幅度之间分别呈中等和弱正相关(r=0.473,P<0.001;r=0.335,P=0.011)。IOL倾斜幅度与α角、IOL屈光度均呈弱正相关(rs=0.272,P=0.003;r=0.285,P=0.002),与ACD、PD、AL呈弱负相关(r=-0.303,P=0.001;r=-0.233,P=0.011;rs=-0.331,P<0.001)。将IOL倾斜幅度与α角、ACD、PD、AL、IOL屈光度纳入多元线性回归分析,采用逐步后退法,建立回归方程为:IOL倾斜幅度=10.503+1.456×α角-0.532×ACD-0.196×AL(R2=0.400;F=8.588,P<0.001)。

结论基于扫频源OCT技术光学生物测量仪可作为IOL倾斜评估的有效方法。IOL倾斜双眼存在镜面对称。

白内障;人工晶状体;生物测量/仪器;倾斜;方向;幅度;扫频源光学相干断层扫描/仪器
ABSTRACT

ObjectiveTo evaluate the feasibility of measurement and characteristics of intraocular lens (IOL) tilt using the swept-source optical coherence tomography (SS-OCT) biometer (IOLMaster 700) and to explore its potentially relevant parameters.

MethodsA cross-sectional study was conducted.Two hundred and forty-two eyes (119 right eyes, 123 left eyes) of 185 patients after phacoemulsification and IOL implantation were included in Beijing Tongren Hospital from July to September 2018.The IOL position, angle κ, angle α, corneal curvature, anterior chamber depth (ACD), pupil diameter (PD), and axial length (AL) were obtained by IOLMaster 700, and the IOL tilt direction and magnitude were calculated.The within-subject standard deviation and intraclass correlation coefficient were used to evaluate the repeatability of three IOL tilt measurements.Binocular symmetry of IOL tilt and the correlation between IOL tilt and different influencing factors were evaluated by Pearson linear correlation analysis or Spearman rank correlation analysis.Influencing factors for IOL tilt were assessed by multiple linear regression analysis.This study adhered to the Declaration of Helsinki.The study protocol was approved by the Ethics Committee of Beijing Tongren Hospital, Capital Medical University (No.TRECKY2018-049). Patients were informed of the purpose and methods of this study and signed the written informed consent.

ResultsThe repeatability of IOL tilt measurements was excellent with a within-subject standard deviation of 5.16° for IOL tilt direction and 0.13° for IOL tilt magnitude.IOL tilt was mirror symmetric in both eyes.The IOL tilt direction in right eyes ranged from -89.60° to 87.53° with a mean of (18.85±29.65)°, and the tilt magnitude ranged from 1.32° to 7.05° with a mean of (4.16±1.26)°.The IOL tilt direction in left eyes ranged from -84.30° to 89.44° with a mean of (21.17±32.38)°, and the tilt magnitude ranged from 0.58° to 7.40° with a mean of (3.80±1.31)°.There were moderate and weak positive correlations of IOL tilt direction and tilt magnitude between both eyes (r=0.473, P<0.001;r=0.335, P=0.011). IOL tilt magnitude was weakly positively correlated with angle α and IOL diopter ( rs=0.272, P=0.003; r=0.285, P=0.002), and was weakly negatively correlated with ACD, PD and AL (r=-0.303, P=0.001; r=-0.233, P=0.011; rs=-0.331, P<0.001). In backward stepwise regression analysis, the regression equation of IOL tilt magnitude, angle α, ACD, PD, AL and IOL diopter in multiple linear regression analysis was as follows: IOL tilt magnitude=10.503+ 1.456×angle α-0.532×ACD-0.196×AL (R2=0.400; F=8.588, P<0.001).

ConclusionsThe SS-OCT biometer can be an effective method to assess IOL tilt.IOL tilt is mirror symmetric between the right eyes and left eyes.

Cataract;Lenses, intraocular;Biometry/instrumentation;Tilt;Direction;Magnitude;Swept-source optical coherence tomography/instrumentation
Yang Wenli, Email: mocdef.3ab61rt_lwgnay
引用本文

赵琦,杨文利,李栋军,等. 基于扫频源OCT技术光学生物测量仪对IOL倾斜测量的可行性评价及倾斜特征分析[J]. 中华实验眼科杂志,2023,41(06):554-560.

DOI:10.3760/cma.j.cn115989-20210311-00167

PERMISSIONS

Request permissions for this article from CCC.

评价本文
*以上评分为匿名评价
白内障摘除联合人工晶状体(intraocular lens,IOL)植入术后IOL处于眼内正确位置是确保术眼最佳视觉质量的主要因素之一。研究发现白内障摘除联合IOL植入术后IOL常发生倾斜[ 1 ],倾斜程度以IOL轴与相应参考轴之间的角度表示。IOL倾斜度超过5°可能对视觉感受带来较为明显的影响[ 2 ],尤其是非球面IOL、复曲面IOL和多焦点IOL植入眼[ 3 , 4 , 5 , 6 , 7 , 8 ]。近年来,基于扫频源光学相干断层扫描(swept source optical coherence tomography,SS-OCT)技术生物测量仪在临床上应用日趋广泛,其可对晶状体进行轴位多角度的B扫描成像,故可用于晶状体倾斜的评估[ 9 , 10 ]。但既往缺乏较大样本针对IOL倾斜测量的可重复性、倾斜相关因素等的研究。本研究拟评估IOLMaster 700对IOL倾斜测量的可重复性,分析IOL的倾斜状态及其影响因素。
试读结束,您可以通过登录机构账户或个人账户后获取全文阅读权限。
参考文献
[1]
Ale JB . Intraocular lens tilt and decentration:a concern for contemporary IOL designs[J]Nepal J Ophthalmol 20113(1)∶6877. DOI: 10.3126/nepjoph.v3i1.4281 .
返回引文位置Google Scholar
百度学术
万方数据
[2]
Korynta J Bok J Cendelin J et al. Computer modeling of visual impairment caused by intraocular lens misalignment[J]J Cataract Refract Surg 199925(1)∶100105. DOI: 10.1016/s0886-3350(99)80019-4 .
返回引文位置Google Scholar
百度学术
万方数据
[3]
Rosales P De Castro A Jiménez-Alfaro I et al. Intraocular lens alignment from purkinje and Scheimpflug imaging[J]Clin Exp Optom 201093(6)∶400408. DOI: 10.1111/j.1444-0938.2010.00514.x .
返回引文位置Google Scholar
百度学术
万方数据
[4]
Montés-Micó R López-Gil N Pérez-Vives C et al. In vitro optical performance of nonrotational symmetric and refractive-diffractive aspheric multifocal intraocular lenses:impact of tilt and decentration[J]J Cataract Refract Surg 201238(9)∶16571663. DOI: 10.1016/j.jcrs.2012.03.040 .
返回引文位置Google Scholar
百度学术
万方数据
[5]
Weikert MP Golla A Wang L Astigmatism induced by intraocular lens tilt evaluated via ray tracing[J]J Cataract Refract Surg 201844(6)∶745749. DOI: 10.1016/j.jcrs.2018.04.035 .
返回引文位置Google Scholar
百度学术
万方数据
[6]
Lawu T Mukai K Matsushima H et al. Effects of decentration and tilt on the optical performance of 6 aspheric intraocular lens designs in a model eye[J]J Cataract Refract Surg 201945(5)∶662668. DOI: 10.1016/j.jcrs.2018.10.049 .
返回引文位置Google Scholar
百度学术
万方数据
[7]
Pérez-Gracia J Varea A Ares J et al. Evaluation of the optical performance for aspheric intraocular lenses in relation with tilt and decenter errors[J/OL]PLoS One 202015(5)∶e0232546[2022-06-13]https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7197786/. DOI: 10.1371/journal.pone.0232546 .
返回引文位置Google Scholar
百度学术
万方数据
[8]
徐雯李茜重视Toric人工晶状体应用误差来源分析和控制[J]中华实验眼科杂志 201836(3)∶165168. DOI: 10.3760/cma.j.issn.2095-0160.2018.03.002 .
返回引文位置Google Scholar
百度学术
万方数据
Xu W Li Q Paying attention to the analysis and control of errors of Toric intraocular lens application[J]Chin J Exp Ophthalmol 201836(3)∶165168. DOI: 10.3760/cma.j.issn.2095-0160.2018.03.002 .
Goto CitationGoogle Scholar
Baidu Scholar
Wanfang Data
[9]
Hirnschall N Buehren T Bajramovic F et al. Prediction of postoperative intraocular lens tilt using swept-source optical coherence tomography[J]J Cataract Refract Surg 201743(6)∶732736. DOI: 10.1016/j.jcrs.2017.01.026 .
返回引文位置Google Scholar
百度学术
万方数据
[10]
Wang L Guimaraes de Souza R Weikert MP et al. Evaluation of crystalline lens and intraocular lens tilt using a swept-source optical coherence tomography biometer[J]J Cataract Refract Surg 201945(1)∶3540. DOI: 10.1016/j.jcrs.2018.08.025 .
返回引文位置Google Scholar
百度学术
万方数据
[11]
Shrout PE . Measurement reliability and agreement in psychiatry[J]Stat Methods Med Res 19987(3)∶301317. DOI: 10.1177/096228029800700306 .
返回引文位置Google Scholar
百度学术
万方数据
[12]
de Castro A Rosales P Marcos S Tilt and decentration of intraocular lenses in vivo from Purkinje and Scheimpflug imaging.Validation study[J]J Cataract Refract Surg 200733(3)∶418429. DOI: 10.1016/j.jcrs.2006.10.054 .
返回引文位置Google Scholar
百度学术
万方数据
[13]
Mura JJ Pavlin CJ Condon GP et al. Ultrasound biomicroscopic analysis of iris-sutured foldable posterior chamber intraocular lenses[J]Am J Ophthalmol 2010149(2)∶245252. DOI: 10.1016/j.ajo.2009.08.022 .
返回引文位置Google Scholar
百度学术
万方数据
[14]
Wang X Dong J Wang X et al. IOL tilt and decentration estimation from 3 dimensional reconstruction of OCT image[J/OL]PLoS One 20138(3)∶e59109[2022-06-19]https://pubmed.ncbi.nlm.nih.gov/23554982/. DOI: 10.1371/journal.pone.0059109 .
返回引文位置Google Scholar
百度学术
万方数据
[15]
Xin C Bian GB Zhang H et al. Optical coherence tomography-based deep learning algorithm for quantification of the location of the intraocular lens[J/OL]Ann Transl Med 20208(14)∶872[2022-06-19]https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7396772/. DOI: 10.21037/atm-20-4706 .
返回引文位置Google Scholar
百度学术
万方数据
[16]
Janunts E Chashchina E Seitz B et al. Reliability of a single light source Purkinjemeter in pseudophakic eyes[J]Optom Vis Sci 201592(8)∶884891. DOI: 10.1097/OPX.0000000000000644 .
返回引文位置Google Scholar
百度学术
万方数据
[17]
Chang DH Waring GO 4th . The subject-fixated coaxially sighted corneal light reflex:a clinical marker for centration of refractive treatments and devices[J]Am J Ophthalmol 2014158(5)∶863874. DOI: 10.1016/j.ajo.2014.06.028 .
返回引文位置Google Scholar
百度学术
万方数据
[18]
Baumeister M Neidhardt B Strobel J et al. Tilt and decentration of three-piece foldable high-refractive silicone and hydrophobic acrylic intraocular lenses with 6-mm optics in an intraindividual comparison[J]Am J Ophthalmol 2005140(6)∶10511058. DOI: 10.1016/j.ajo.2005.07.026 .
返回引文位置Google Scholar
百度学术
万方数据
[19]
Mutlu FM Erdurman C Sobaci G et al. Comparison of tilt and decentration of 1-piece and 3-piece hydrophobic acrylic intraocular lenses[J]J Cataract Refract Surg 200531(2)∶343347. DOI: 10.1016/j.jcrs.2004.06.022 .
返回引文位置Google Scholar
百度学术
万方数据
[20]
Findl O Hirnschall N Nishi Y et al. Capsular bag performance of a hydrophobic acrylic 1-piece intraocular lens[J]J Cataract Refract Surg 201541(1)∶9097. DOI: 10.1016/j.jcrs.2014.04.030 .
返回引文位置Google Scholar
百度学术
万方数据
[21]
Zhang F Zhang J Li W et al. Correlative comparison of three ocular axes to tilt and decentration of intraocular lens and their effects on visual acuity[J]Ophthalmic Res 202063(2)∶165173. DOI: 10.1159/000504716 .
返回引文位置Google Scholar
百度学术
万方数据
[22]
Nishi Y Hirnschall N Crnej A et al. Reproducibility of intraocular lens decentration and tilt measurement using a clinical Purkinje meter[J]J Cataract Refract Surg 201036(9)∶15291535. DOI: 10.1016/j.jcrs.2010.03.043 .
返回引文位置Google Scholar
百度学术
万方数据
[23]
刘春雷黄芳俞阿勇应用三维眼前节分析仪对单片式折叠型人工晶状体在眼内位置分布规律的研究[J]中华实验眼科杂志 201129(10)∶913917. DOI: 10.3760/cma.j.issn.2095-0160.2011.10.011 .
返回引文位置Google Scholar
百度学术
万方数据
Liu CL Huang F Yu AY et al. Analysis of position of one-piece soft intraocular lens after implantation of intraocular lens with Pentacam[J]Chin J Exp Ophthalmol 201129(10)∶913917. DOI: 10.3760/cma.j.issn.2095-0160.2011.10.011 .
Goto CitationGoogle Scholar
Baidu Scholar
Wanfang Data
[24]
Kimura S Morizane Y Shiode Y et al. Assessment of tilt and decentration of crystalline lens and intraocular lens relative to the corneal topographic axis using anterior segment optical coherence tomography[J/OL]PLoS One 201712(9)∶e0184066[2022-06-20]https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5581187/. DOI: 10.1371/journal.pone.0184066 .
返回引文位置Google Scholar
百度学术
万方数据
[25]
Sato T Shibata S Yoshida M et al. Short-term dynamics after single- and three-piece acrylic intraocular lens implantation:a swept-source anterior segment optical coherence tomography study[J/OL]Sci Rep 20188(1)∶10230[2022-06-23]https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6035277/. DOI: 10.1038/s41598-018-28609-1 .
返回引文位置Google Scholar
百度学术
万方数据
[26]
Harrer A Hirnschall N Tabernero J et al. Variability in angle κ and its influence on higher-order aberrations in pseudophakic eyes[J]J Cataract Refract Surg 201743(8)∶10151019. DOI: 10.1016/j.jcrs.2017.05.028 .
返回引文位置Google Scholar
百度学术
万方数据
[27]
Ang GS Duncan L Atta HR . Ultrasound biomicroscopic study of the stability of intraocular lens implants after phacoemulsification cataract surgery[J]Acta Ophthalmol 201290(2)∶168172. DOI: 10.1111/j.1755-3768.2010.01880.x .
返回引文位置Google Scholar
百度学术
万方数据
[28]
章玉群徐巍华田琳璐年龄相关性白内障超声乳化吸出联合人工晶状体植入术后Pentacam眼前节分析[J]国际眼科杂志 201111(2)∶235239. DOI: 10.3969/j.issn.1672-5123.2011.02.12 .
返回引文位置Google Scholar
百度学术
万方数据
Zhang YQ Xu WH Tian LL et al. Analysis of anterior segment of age-related cataract after phacoemulsification combinded with intraocular lens implantation by Pentacam[J]Int Eye Sci 201111(2)∶235239. DOI: 10.3969/j.issn.1672-5123.2011.02.12 .
Goto CitationGoogle Scholar
Baidu Scholar
Wanfang Data
[29]
Chen X Gu X Wang W et al. Characteristics and factors associated with intraocular lens tilt and decentration after cataract surgery[J]J Cataract Refract Surg 202046(8)∶11261131. DOI: 10.1097/j.jcrs.0000000000000219 .
返回引文位置Google Scholar
百度学术
万方数据
[30]
Zhu X He W Zhang Y et al. Inferior decentration of multifocal intraocular lenses in myopic eyes[J]Am J Ophthalmol 201818818. DOI: 10.1016/j.ajo.2018.01.007 .
返回引文位置Google Scholar
百度学术
万方数据
[31]
Kránitz K Miháltz K Sándor GL et al. Intraocular lens tilt and decentration measured by Scheimpflug camera following manual or femtosecond laser-created continuous circular capsulotomy[J]J Refract Surg 201228(4)∶259263. DOI: 10.3928/1081597X-20120309-01 .
返回引文位置Google Scholar
百度学术
万方数据
[32]
Artal P Guirao A Berrio E et al. Compensation of corneal aberrations by the internal optics in the human eye[J]J Vis 20011(1)∶18. DOI: 10.1167/1.1.1 .
返回引文位置Google Scholar
百度学术
万方数据
备注信息
A
杨文利,Email:mocdef.3ab61rt_lwgnay
B

赵琦:参与选题、设计及实施研究、采集数据、撰写文章;杨文利、鲜军舫:参与设计研究、文章知识性内容的审阅和智力性内容的修改及定稿;李栋军、王子杨、陈伟、李逸丰、崔蕊、沈琳:实施研究、采集数据;李蕾:分析/解释数据

C
所有作者均声明不存在利益冲突
D
北京市医院管理局临床医学发展专项项目 (ZYLX201704)
北京市卫生系统高层次人才培养计划学科带头人项目 (2014-2-005)
评论 (0条)
注册
登录
时间排序
暂无评论,发表第一条评论抢沙发
MedAI助手(体验版)
文档即答
智问智答
机器翻译
回答内容由人工智能生成,我社无法保证其准确性和完整性,该生成内容不代表我们的态度或观点,仅供参考。
生成快照
文献快照

你好,我可以帮助您更好的了解本文,请向我提问您关注的问题。

0/2000

《中华医学会杂志社用户协议》 | 《隐私政策》

《SparkDesk 用户协议》 | 《SparkDesk 隐私政策》

网信算备340104764864601230055号 | 网信算备340104726288401230013号

技术支持:

历史对话
本文全部
还没有聊天记录
设置
模式
纯净模式沉浸模式
字号