临床研究
ENGLISH ABSTRACT
Scansys、Pentacam、CASIA和Arcscan测量近视眼ICL植入术后拱高比较
张波
王浩
庞辰久
杜文文
孙早荷
李金
顾宇伟
王树林
范棋
作者及单位信息
·
DOI: 10.3760/cma.j.cn115989-20221023-00492
Comparison of vault measurements by Scansys, Pentacam, CASIA and Arcscan after ICL implantation
Zhang Bo
Wang Hao
Pang Chenjiu
Du Wenwen
Sun Zaohe
Li Jin
Gu Yuwei
Wang Shulin
Fan Qi
Authors Info & Affiliations
Zhang Bo
Henan Provincial People's Hospital, Henan Eye Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan University, Zhengzhou 450003, China
Wang Hao
Henan Provincial People's Hospital, Henan Eye Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan University, Zhengzhou 450003, China
Pang Chenjiu
Henan Provincial People's Hospital, Henan Eye Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan University, Zhengzhou 450003, China
Du Wenwen
Henan Provincial People's Hospital, Henan Eye Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan University, Zhengzhou 450003, China
Sun Zaohe
Henan Provincial People's Hospital, Henan Eye Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan University, Zhengzhou 450003, China
Li Jin
Henan Provincial People's Hospital, Henan Eye Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan University, Zhengzhou 450003, China
Gu Yuwei
Henan Provincial People's Hospital, Henan Eye Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan University, Zhengzhou 450003, China
Wang Shulin
Henan Provincial People's Hospital, Henan Eye Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan University, Zhengzhou 450003, China
Fan Qi
Henan Provincial People's Hospital, Henan Eye Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan University, Zhengzhou 450003, China
·
DOI: 10.3760/cma.j.cn115989-20221023-00492
426
98
0
1
4
0
PDF下载
APP内阅读
摘要

目的评估国产三维眼前节分析系统Scansys、Pentacam三维眼前节全景分析仪、频域眼前节光学相干断层扫描仪(CASIA SS-1000)和新型超高频数字超声扫描系统(Arcscan Insight100)测量近视眼有晶状体眼后房型人工晶状体(ICL)植入术后中央拱高的差异及一致性。

方法采用诊断试验研究方法,纳入2019年6—12月于河南省人民医院眼科行ICL V4c植入术的近视患者56例56眼,术后同时应用Scansys、Pentacam、CASIA和Arcsan测量中央拱高。比较4种仪器拱高测量值的差异;4种仪器拱高测量值的相关性采用Pearson相关性分析,一致性比较采用Bland-Altman法分析。

结果Scansys、Pentacam、CASIA和Arcscan拱高测量值分别为(481.8±191.6)、(476.4±190.6)、(619.3±207.5)和(534.0±221.2)μm,总体比较差异有统计学意义( F=143.301, P<0.001),其中Scansys和Pentacam拱高测量值明显低于CASIA和Arcscan,Arcscan拱高测量值低于CASIA,差异均有统计学意义(均 P<0.001)。Arcscan-CASIA、Arcscan-Pentacam、Arcscan-Scansys、CASIA-Pentacam、CASIA-Scansys、Pentacam-Scansys拱高测量值均呈强正相关( r=0.982、0.933、0.931、0.942、0.941、0.989,均 P<0.001)。Scansys、Pentacam、CASIA和Arcscan拱高测量值组内相关系数分别为0.985、0.975、0.998、0.992;CASIA和Arcscan、CASIA和Scansys、CASIA和Pantacam、Arcscan和Scansys、Arcscan和Pantacam拱高测量值差值95%一致性界限分别为-170~0、0~280、0~280、-110~210、-100~220 μm,差值的绝对值最大值超出临床的可接受范围,一致性欠佳;Scansys和Pentacam拱高测量值差值95%一致性界限为-60~50 μm,一致性良好。

结论4种仪器测量近视眼ICL V4c植入术后拱高重复性均良好,其中,Scansys和Pentacam的拱高测量值偏小,两者一致性良好,结果可以互相替代;而CASIA拱高测量值最大,其次为Arcscan,彼此差异较大,测量结果不能相互替代,在临床工作中应结合实际情况综合分析。

近视;有晶状体眼人工晶状体;拱高;生物测量;超声,数字式,超高频;光学相干成像,前节,频域;三维眼前节全景分析仪;一致性
ABSTRACT

ObjectiveTo evaluate the consistency of the Chinese three-dimensional anterior visual field analysis system (Scansys), the anterior segment analyzer (Pentacam), the frequency-domain anterior segment optical coherence tomography system (CASIA SS-1000), and a new ultra-high frequency digital ultrasound scanning system (Arcscan Insight100) to measure central vault after implantable collamer lens (ICL) implantation in myopic eyes with crystalline lenses.

MethodsA diagnostic test study was conducted.Fifty-six myopic patients (56 eyes) who underwent ICL V4c implantation from June to December 2019 were included.Scansys, Pentacam, CASIA and Arcscan were used to measure the central vault after surgery.The vault measurements were compared.Correlations between the measurements of the four instruments were analyzed using Pearson correlation analysis, and consistency comparisons were analyzed using the Bland-Altman method.This study adhered to the Declaration of Helsinki.The study protocol was approved by the Ethics Committee of Henan Eye Hospital (No.HNEECKY-2021[13]). Written informed consent was obtained from each subject.

ResultsThe central vault measurements by Scansys, Pentacam, CASIA and Arcscan were (481.8±191.6), (476.4±190.6), (619.3±207.5) and (534.0±221.2)μm, respectively, with a statistically significant overall difference ( F=143.301, P<0.001). The vault measurements by Scansys and Pentacam were significantly lower than CASIA and Arcscan, and Arcscan was lower than CASIA, with statistically significant differences (all at P<0.001). There were strong positive correlations in vault measurements between Arcscan and CASIA, Arcscan and Pentacam, Arcscan and Scansys, CASIA and Pentacam, CASIA and Scansys, Pentacam and Scansys ( r=0.982, 0.933, 0.931, 0.942, 0.941, 0.989; all at P<0.001). Intraclass correlation coefficients of vault measurements by Scansys, Pentacam, CASIA and Arcscan were 0.985, 0.975, 0.998, 0.992, respectively.The 95% limits of agreement of vault measurements differences were -170 to 0, 0 to 280, 0 to 280, -110 to 210, -100 to 220 μm, between CASIA and Arcscan, CASIA and Scansys, CASIA and Pentacam, Arcscan and Scansys, Arcscan and Pentacam, respectively, and the maximum absolute value of the difference was beyond the clinically acceptable range, showing poor agreement.The 95% limits of agreement of vault measurement difference was -60 to 50 μm between Scansys and Pentacam, showing a good agreement.

ConclusionsThe repeatability of the vault after ICL V4c implantation in myopic eyes measured by the four instruments is good.Among them, the vault measurements of Scansys and Pentacam are smaller, showing good consistency, and their results could be substituted for each other.The measurement of CASIA is the largest, followed by Arcscan, which have a large difference from each other, and their results can not be substituted for each other, which should be comprehensively analyzed with the actual situation in clinical work.

Myopia;Implantable collamer lens;Vault;Bio-measurement;Ultrasonic, digital, UHF;Optical coherence tomography, anterior segment, frequency domain;Three-dimensional anterior segment panoramic analyzer;Agreement
Pang Chenjiu, Email: mocdef.3ab61999jcgnap
引用本文

张波,王浩,庞辰久,等. Scansys、Pentacam、CASIA和Arcscan测量近视眼ICL植入术后拱高比较[J]. 中华实验眼科杂志,2024,42(04):354-360.

DOI:10.3760/cma.j.cn115989-20221023-00492

PERMISSIONS

Request permissions for this article from CCC.

评价本文
*以上评分为匿名评价
有晶状体眼后房型人工晶状体(implantable collamer lens,ICL)植入术具有矫正范围广、安全、稳定、可逆等优势 [ 1 , 2 , 3 , 4 ],目前不仅用于高度近视、超高度近视矫正,而且逐渐应用于低中度近视矫正。ICL植入术后眼前节形态,如前房深度(anterior chamber depth,ACD)、前房角、ICL位置等是评估ICL稳定性及安全性的重要指标,其中拱高(即ICL后表面与透明晶状体前表面间距)是临床关注的焦点之一,也是术后决定是否需要调换ICL的重要依据。目前,眼科临床常用的综合检查测量眼前节结构的仪器有Scheimpflug相机原理成像的三维眼前节全景分析仪(如Pentacam)、眼前节光学相干断层扫描(anterior segment optical coherence tomography,AS-OCT)和超声生物显微镜(ultrasound biomicroscope,UBM),应用中各有优点和局限性。近期出现的新型超高频数字超声扫描系统(Arcscan Insight100)通过对眼球进行弧形全自动扫描,快速生成特定子午线上的眼前节影像图,可对眼前节的综合解剖结构进行探查和精确测量,而且无需表面麻醉及卧位,眼部不用放置浴杯,受检查者操作主观因素影响小。目前,Arcscan Insight100应用于ICL植入术后眼前节结构的观察及与其他仪器同时测量对比的相关研究报道国内外尚缺乏。本研究旨在对比Scheimpflug相机原理成像的三维眼前节分析系统Scansys、Pentacam HR、频域AS-OCT和Arcscan Insight100测量近视眼ICL V4c植入手术后中央拱高的差异及一致性,以期为临床应用提供参考。
试读结束,您可以通过登录机构账户或个人账户后获取全文阅读权限。
参考文献
[1]
Papa-Vettorazzi MR , Güell JL , Cruz-Rodriguez JB et al. Long-term efficacy and safety profiles after posterior chamber phakic intraocular lens implantation in eyes with more than 10 years of follow-up[J]J Cataract Refract Surg 202248(7)∶813-818. DOI: 10.1097/j.jcrs.0000000000000848 .
返回引文位置Google Scholar
百度学术
万方数据
[2]
Choi JH , Lim DH , Nam SW et al. Ten-year clinical outcomes after implantation of a posterior chamber phakic intraocular lens for myopia[J]J Cataract Refract Surg 201945(11)∶1555-1561. DOI: 10.1016/j.jcrs.2019.06.015 .
返回引文位置Google Scholar
百度学术
万方数据
[3]
Qin Q , Wu Z , Bao L et al. Evaluation of visual quality after EVO-ICL implantation for hypermyopia:an observational study[J/OL]Medicine (Baltimore) 201998(44)∶e17677[2023-10-09]https://pubmed.ncbi.nlm.nih.gov/31689784/. DOI: 10.1097/MD.0000000000017677 .
返回引文位置Google Scholar
百度学术
万方数据
[4]
Fernández-Vega-Cueto L , Lisa C , Esteve-Taboada JJ et al. Implantable collamer lens with central hole:3-year follow-up[J]Clin Ophthalmol 2018122015-2029. DOI: 10.2147/OPTH.S171576 .
返回引文位置Google Scholar
百度学术
万方数据
[5]
中华医学会眼科学分会眼视光学组中国有晶状体眼后房型人工晶状体植入术专家共识(2019年)[J]中华眼科杂志 201955(9)∶652-657. DOI: 10.3760/cma.j.issn.0412-4081.2019.09.005 .
返回引文位置Google Scholar
百度学术
万方数据
[6]
Kamiya K , Shimizu K , Komatsu M Factors affecting vaulting after implantable collamer lens implantation[J]J Refract Surg 200925(3)∶259-264. DOI: 10.3928/1081597X-20090301-04 .
返回引文位置Google Scholar
百度学术
万方数据
[7]
Lege BA , Haigis W , Neuhann TF et al. Age-related behavior of posterior chamber lenses in myopic phakic eyes during accommodation measured by anterior segment partial coherence interferometry[J]J Cataract Refract Surg 200632(6)∶999-1006. DOI: 10.1016/j.jcrs.2006.02.018 .
返回引文位置Google Scholar
百度学术
万方数据
[8]
Zhu QJ , Chen WJ , Zhu WJ et al. Short-term changes in and preoperative factors affecting vaulting after posterior chamber phakic implantable collamer lens implantation[J/OL]BMC Ophthalmol 202121(1)∶199[2023-10-10]https://pubmed.ncbi.nlm.nih.gov/33957891/. DOI: 10.1186/s12886-021-01963-x .
返回引文位置Google Scholar
百度学术
万方数据
[9]
Trancón AS , Manito SC , Sierra OT et al. Determining vault size in implantable collamer lenses:preoperative anatomy and lens parameters[J]J Cataract Refract Surg 202046(5)∶728-736. DOI: 10.1097/j.jcrs.0000000000000146 .
返回引文位置Google Scholar
百度学术
万方数据
[10]
Lee DH , Choi SH , Chung ES et al. Correlation between preoperative biometry and posterior chamber phakic Visian implantable collamer lens vaulting[J]Ophthalmology 2012119(2)∶272-277. DOI: 10.1016/j.ophtha.2011.07.047 .
返回引文位置Google Scholar
百度学术
万方数据
[11]
Kato S , Shimizu K , Igarashi A Vault changes caused by light-induced pupil constriction and accommodation in eyes with an implantable collamer lens[J]Cornea 201938(2)∶217-220. DOI: 10.1097/ICO.0000000000001785 .
返回引文位置Google Scholar
百度学术
万方数据
[12]
Lu Y , Yang N , Li X et al. Four-year follow-up of the changes in anterior segment after phakic collamer lens implantation[J]Am J Ophthalmol 2017178140-149. DOI: 10.1016/j.ajo.2017.03.020 .
返回引文位置Google Scholar
百度学术
万方数据
[13]
Moya T , Javaloy J , Montés-Micó R et al. Implantable collamer lens for myopia:assessment 12 years after implantation[J]J Refract Surg 201531(8)∶548-556. DOI: 10.3928/1081597X-20150727-05 .
返回引文位置Google Scholar
百度学术
万方数据
[14]
Schmidinger G , Lackner B , Pieh S et al. Long-term changes in posterior chamber phakic intraocular collamer lens vaulting in myopic patients[J]Ophthalmology 2010117(8)∶1506-1511. DOI: 10.1016/j.ophtha.2009.12.013 .
返回引文位置Google Scholar
百度学术
万方数据
[15]
董晶高晓唯胡裕坤Pentacam与前节OCT测量有晶状体眼后房型人工晶状体植入术后拱高的比较[J]中华眼视光学与视觉科学杂志 201618(12)∶724-728. DOI: 10.3760/cma.j.issn.1674-845X.2016.12.005 .
返回引文位置Google Scholar
百度学术
万方数据
Dong J , Gao XW , Hu YK et al. Comparison of vault measurements by Pentacam and anterior segment OCT after Collamer lens implantation[J]Chin J Optom Ophthalmol Vis Sci 201618(12)∶724-728. DOI: 10.3760/cma.j.issn.1674-845X.2016.12.005 .
Goto CitationGoogle Scholar
Baidu Scholar
Wanfang Data
[16]
薛艳珍周奇志有晶状体眼后房型人工晶状体植入术后UBM与前节OCT测量拱高一致性比较[J]中华实验眼科杂志 201937(4)∶287-291. DOI: 10.3760/cma.j.issn.2095-0160.2019.04.010 .
返回引文位置Google Scholar
百度学术
万方数据
Xue YZ , Zhou QZ . Agreement of central vault measurements between UBM and anterior segment OCT in posterior chamber intraocular lens implanted eyes[J]Chin J Exp Ophthalmol 201937(4)∶287-291. DOI: 10.3760/cma.j.issn.2095-0160.2019.04.010 .
Goto CitationGoogle Scholar
Baidu Scholar
Wanfang Data
[17]
Wan T , Yin H , Yang Y et al. Comparative study of anterior segment measurements using 3 different instruments in myopic patients after ICL implantation[J/OL]BMC Ophthalmol 201919(1)∶182[2023-10-10]https://pubmed.ncbi.nlm.nih.gov/31409385/. DOI: 10.1186/s12886-019-1194-y .
返回引文位置Google Scholar
百度学术
万方数据
[18]
Reinstein DZ , Vida RS , Archer TJ . Visual outcomes,footplate position and vault achieved with the Visian implantable collamer lens for myopic astigmatism[J]Clin Ophthalmol 2021154485-4497. DOI: 10.2147/OPTH.S330879 .
返回引文位置Google Scholar
百度学术
万方数据
[19]
Lindland A , Heger H , Kugelberg M et al. Changes in vaulting of myopic and toric implantable collamer lenses in different lighting conditions[J]Acta Ophthalmol 201290(8)∶788-791. DOI: 10.1111/j.1755-3768.2011.02224.x .
返回引文位置Google Scholar
百度学术
万方数据
[20]
Srirampur A , Pesala V , Mansoori T et al. Effect of pupil size on posterior chamber phakic intraocular lens vault measurements[J]Indian J Ophthalmol 202169(9)∶2289-2292. DOI: 10.4103/ijo.IJO_3429_20 .
返回引文位置Google Scholar
百度学术
万方数据
备注信息
A
庞辰久,Email: mocdef.3ab61999jcgnap
B

张波:酝酿和设计试验、实施研究、采集数据、分析/解释数据、文章撰写;王浩、杜文文、孙早荷、李金、顾宇伟、王树林、范棋:采集数据、分析/解释数据、对文章知识性内容作批评性审阅;庞辰久:酝酿和设计试验、指导试验、分析/解释数据、对文章知识性内容作批评性审阅及定稿

C
所有作者均声明不存在利益冲突
D
河南省科技厅科技攻关联合基金 (232103810052)
评论 (0条)
注册
登录
时间排序
暂无评论,发表第一条评论抢沙发
MedAI助手(体验版)
文档即答
智问智答
机器翻译
回答内容由人工智能生成,我社无法保证其准确性和完整性,该生成内容不代表我们的态度或观点,仅供参考。
生成快照
文献快照

你好,我可以帮助您更好的了解本文,请向我提问您关注的问题。

0/2000

《中华医学会杂志社用户协议》 | 《隐私政策》

《SparkDesk 用户协议》 | 《SparkDesk 隐私政策》

网信算备340104764864601230055号 | 网信算备340104726288401230013号

技术支持:

历史对话
本文全部
还没有聊天记录
设置
模式
纯净模式沉浸模式
字号