两种扫描杆辅助下全牙弓种植固定修复口内扫描精度的比较
富晓娇
蔡征真
史俊宇
赖红昌
作者及单位信息
·
DOI: 10.3760/cma.j.cn112144-20240817-00315
Comparison of the accuracy of two types of scanning bodies for intraoral scanning in complete-arch implant-supported fixed restoration
Fu Xiaojiao
Cai Zhengzhen
Shi Junyu
Lai Hongchang
Authors Info & Affiliations
Fu Xiaojiao
Department of Oral and Maxillofacial Implantology, Shanghai Ninth People′s Hospital, Shanghai Jiao Tong University School of Medicine & College of Stomatology, Shanghai Jiao Tong University & National Center for Stomatology & National Clinical Research Center for Oral Diseases & Shanghai Key Laboratory of Stomatology & Shanghai Research Institute of Stomatology, Shanghai 200011, China
Cai Zhengzhen
Department of Oral and Maxillofacial Implantology, Shanghai Ninth People′s Hospital, Shanghai Jiao Tong University School of Medicine & College of Stomatology, Shanghai Jiao Tong University & National Center for Stomatology & National Clinical Research Center for Oral Diseases & Shanghai Key Laboratory of Stomatology & Shanghai Research Institute of Stomatology, Shanghai 200011, China
Shi Junyu
Department of Oral and Maxillofacial Implantology, Shanghai Ninth People′s Hospital, Shanghai Jiao Tong University School of Medicine & College of Stomatology, Shanghai Jiao Tong University & National Center for Stomatology & National Clinical Research Center for Oral Diseases & Shanghai Key Laboratory of Stomatology & Shanghai Research Institute of Stomatology, Shanghai 200011, China
Lai Hongchang
Department of Oral and Maxillofacial Implantology, Shanghai Ninth People′s Hospital, Shanghai Jiao Tong University School of Medicine & College of Stomatology, Shanghai Jiao Tong University & National Center for Stomatology & National Clinical Research Center for Oral Diseases & Shanghai Key Laboratory of Stomatology & Shanghai Research Institute of Stomatology, Shanghai 200011, China
·
DOI: 10.3760/cma.j.cn112144-20240817-00315
154
34
0
0
0
0
PDF下载
APP内阅读
摘要

目的探讨全牙弓种植固定修复过程中常规扫描杆、伞形扫描杆、伞形扫描杆结合辅助杆辅助下口内扫描的精度差异,以期为伞形扫描杆的临床应用提供参考。

方法自主研发1种新型伞形扫描杆以及配套辅助杆。制作带有6个平行种植体复合基台替代体的上颌超硬石膏模型1个。于石膏模型上安装常规扫描杆,采用牙颌模型扫描仪进行模型扫描作为参考数据。将常规扫描杆、伞形扫描杆、伞形扫描杆结合辅助杆分别安装于石膏模型上,由1名主治医师采用口内扫描仪分别重复扫描10次,作为实验组数据(常规扫描杆组、伞形扫描杆组、伞形扫描杆结合辅助杆组)。将实验组数据与参考数据生成的虚拟基台模型进行最佳拟合配准,测量并计算分别代表正确度和精密度的均方根误差(RMSE)值及各实验组基台间距离偏差和角度偏差,以及扫描时间。采用重复测量的方差分析和广义估计方程模型进行统计学分析。

结果伞形扫描杆组和伞形扫描杆结合辅助杆组正确度RMSE值[(48.0±12.6)和(45.9±13.4)μm]和距离偏差[(64.5±60.2)和(63.8±54.4)μm]均显著优于常规扫描杆组[(81.9±23.9)和(90.0±85.2)μm](均 P<0.05),伞形扫描杆组和伞形扫描杆结合辅助杆组正确度RMSE值和距离偏差差异均无统计学意义(均 P>0.05)。3组精密度RMSE值和角度偏差差异均无统计学意义(均 P>0.05)。伞形扫描杆组和伞形扫描杆结合辅助杆组扫描时间[(54.3±11.8)和(35.8±10.1)s]均显著小于常规扫描杆组[(108.7±38.9)s](均 P<0.05)。

结论相比常规扫描杆,新型伞形扫描杆在针对全牙弓种植固定修复进行口内扫描时,具有更高的精准度与效率。

牙种植;颌,无牙;口内扫描;精度;数字化口腔医学
ABSTRACT

ObjectiveTo compare the accuracy of intraoral scanning for complete-arch implant-supported fixed prosthesis using umbrella scan bodies (USB) and conventional scan bodies (CSB), providing a reference for the clinical application of umbrella-shaped scanning bodies.

MethodsA new type of umbrella-shaped scanning body and its matching auxiliary bar were independently developed. A maxillary type Ⅳ dental stone model with six parallel implant abutment analogs was fabricated. Conventional scanning bodies were installed on the model, and a laboratory scanner was used to scan the model as reference data. The CSB, USB, and USB combined with an auxiliary bar (U+SB) were installed on the model, respectively. A single attending physician performed intraoral scanning 10 times for each group using an intraoral scanner, serving as test group data (CSB, USB, U+SB). The test data were best-fit aligned with the virtual abutment models generated from the reference data. The trueness and precision of root-mean-square error (RMSE) values, inter-abutment distance deviations, angular deviations, and scanning time were measured and calculated. Repeated measures ANOVA and generalized estimating equation models were used for statistical analysis.

ResultsThe trueness of RMSE values [(48.0±12.6) and (45.9±13.4) μm] and distance deviations [(64.5±60.2) and (63.8±54.4) μm] of the USB and U+SB groups were significantly better than those of the CSB group [(81.9±23.9) and (90.0±85.2) μm] (all P<0.05). There was no significant difference in trueness of RMSE values and distance deviations between the USB group and U+SB group (all P>0.05). There were no significant differences in the precision of RMSE values and angular deviations among the three groups (all P>0.05). The scanning time of the USB group and U+SB group [(54.3±11.8) and (35.8±10.1) s] was significantly shorter than that of CSB group [(108.7±38.9) s] (all P<0.05).

ConclusionsCompared with conventional scanning bodies, the new umbrella-shaped scanning body demonstrates higher accuracy and efficiency for intraoral scanning impressions in complete-arch implant-supported fixed prosthesis.

Dental implantation;Jaw, edentulous;Intraoral scanning;Accuracy;Digital dentistry
Lai Hongchang, Email: mocdef.6ab21ialgnahcgnoh, Tel: 0086-21-53315299
引用本文

富晓娇,蔡征真,史俊宇,等. 两种扫描杆辅助下全牙弓种植固定修复口内扫描精度的比较[J]. 中华口腔医学杂志,2025,60(03):267-272.

DOI:10.3760/cma.j.cn112144-20240817-00315

PERMISSIONS

Request permissions for this article from CCC.

评价本文
*以上评分为匿名评价
通过口内扫描获得数字化印模在口腔种植领域应用广泛,单牙和短跨度种植时数字化印模的精度可满足临床应用要求 1 , 2。然而,通过口内扫描获得全牙弓种植固定修复数字化印模仍是临床难点 3。主要因为口内扫描的原理是通过深度相机连续获取多帧图像,利用算法进行拼接,拼接的依据是相邻帧图像间共同的解剖形态信息 4。然而,全牙弓种植固定修复患者口内扫描区域解剖标志相对较少,可增加后续图像配准拼接难度 5 , 6,进而引入较大的误差。此外,口内可移动的软组织在扫描过程中可能因呼吸、吞咽或口腔肌肉的活动而发生位移,从而干扰扫描设备精准捕捉图像 7 , 8 , 9。全牙弓扫描范围广,覆盖整个牙弓,在此扫描跨度下,即使每个小区域的扫描误差较小,但累积的总体误差,可对最终的图像配准拼接结果产生不利影响。因此,全牙弓种植固定修复的口内扫描面临解剖标志稀缺、软组织移动干扰以及扫描跨度大等多重问题,这些问题共同影响口内扫描的准确性。
已有研究显示,通过增加扫描杆等辅助装置,于扫描杆上安装延伸结构,可提高全牙弓种植固定修复口内扫描精度 10 , 11。本课题组一项体外研究对比不同扫描仪在采用扫描杆辅助装置时的精度,结果显示,对于不同的扫描仪,扫描杆辅助装置均可显著提高扫描精度;且临床研究显示,口内扫描结合扫描杆辅助装置与口外摄影测量的距离偏差差异无统计学意义 12。然而,现有的全牙弓种植固定修复的扫描杆大多是在常规种植扫描杆的基础上增加辅助装置 1113 , 14 , 15 , 16,个别研究针对全牙弓种植口内扫描应用特定扫描杆 17。本课题组结合口内扫描数字化印模的扫描难点,自主设计研发1种新型伞形扫描杆及配套辅助杆,本项研究探讨全牙弓种植固定修复过程中常规扫描杆、伞形扫描杆、伞形扫描杆结合辅助杆辅助下口内扫描的精度差异,以期为伞形扫描杆的临床应用提供参考。
试读结束,您可以通过登录机构账户或个人账户后获取全文阅读权限。
参考文献
[1]
Nejatidanesh F , Shakibamehr AH , Savabi O . Comparison of marginal and internal adaptation of CAD/CAM and conventional cement retained implant-supported single crowns[J]. Implant Dent, 2016,25(1):103-108. DOI: 10.1097/ID.0000000000000346 .
返回引文位置Google Scholar
百度学术
万方数据
[2]
Schmidt A , Wöstmann B , Schlenz MA . Accuracy of digital implant impressions in clinical studies: a systematic review[J]. Clin Oral Implants Res, 2022,33(6):573-585. DOI: 10.1111/clr.13951 .
返回引文位置Google Scholar
百度学术
万方数据
[3]
Papaspyridakos P , Vazouras K , Chen YW ,et al. Digital vs conventional implant impressions: a systematic review and meta-analysis[J]. J Prosthodont, 2020,29(8):660-678. DOI: 10.1111/jopr.13211 .
返回引文位置Google Scholar
百度学术
万方数据
[4]
Ko MS , Park SC . Alignment of dental depth images from an intraoral scanner[J]. Comput Aided Des Appl, 2018,15(6):863-871. DOI: 10.1080/16864360.2018.1486962 .
返回引文位置Google Scholar
百度学术
万方数据
[5]
Cai ZZ , Li X , Wu XY ,et al. Does intra-oral scan improve the impression accuracy of full-arch implant-supported prostheses: a systematic review and meta-analysis[J]. Clin Implant Dent Relat Res, 2024,26(5):847-861. DOI: 10.1111/cid.13321 .
返回引文位置Google Scholar
百度学术
万方数据
[6]
Zhang YJ , Qiao SC , Qian SJ ,et al. Influence of different factors on the accuracy of digital impressions of multiple implants: an in vitro study[J]. Int J Oral Maxillofac Implants, 2021,36(3):442-449. DOI: 10.11607/jomi.8532 .
返回引文位置Google Scholar
百度学术
万方数据
[7]
Alikhasi M , Siadat H , Nasirpour A ,et al. Three-dimensional accuracy of digital impression versus conventional method: effect of implant angulation and connection type[J]. Int J Dent, 2018,2018:3761750. DOI: 10.1155/2018/3761750 .
返回引文位置Google Scholar
百度学术
万方数据
[8]
Albayrak B , Sukotjo C , Wee AG ,et al. Three-dimensional accuracy of conventional versus digital complete arch implant impressions[J]. J Prosthodont, 2021,30(2):163-170. DOI: 10.1111/jopr.13264 .
返回引文位置Google Scholar
百度学术
万方数据
[9]
Di Fiore A , Meneghello R , Graiff L ,et al. Full arch digital scanning systems performances for implant-supported fixed dental prostheses: a comparative study of 8 intraoral scanners[J]. J Prosthodont Res, 2019,63(4):396-403. DOI: 10.1016/j.jpor.2019.04.002 .
返回引文位置Google Scholar
百度学术
万方数据
[10]
Kernen FR , Recca M , Vach K ,et al. In vitro scanning accuracy using different aids for multiple implants in the edentulous arch[J]. Clin Oral Implants Res, 2022,33(10):1010-1020. DOI: 10.1111/clr.13982 .
返回引文位置Google Scholar
百度学术
万方数据
[11]
Ahmed WM , Verhaeghe TV , McCullagh A . Maxillary complete-arch implant-supported restoration: a digital scanning and maxillomandibular relationship workflow[J]. J Prosthet Dent, 2021,125(2):216-220. DOI: 10.1016/j.prosdent.2020.01.010 .
返回引文位置Google Scholar
百度学术
万方数据
[12]
Fu XJ , Liu M , Liu BL ,et al. Accuracy of intraoral scan with prefabricated aids and stereophotogrammetry compared with open tray impressions for complete-arch implant-supported prosthesis: a clinical study[J]. Clin Oral Implants Res, 2024,35(8):830-840. DOI: 10.1111/clr.14183 .
返回引文位置Google Scholar
百度学术
万方数据
[13]
Lawand G , Ismail Y , Revilla-León M ,et al. Effect of implant scan body geometric modifications on the trueness and scanning time of complete arch intraoral implant digital scans: an in vitro study[J]. J Prosthet Dent, 2024,131(6):1189-1197. DOI: 10.1016/j.prosdent.2022.06.004 .
返回引文位置Google Scholar
百度学术
万方数据
[14]
Kernen F , Schlager S , Seidel Alvarez V ,et al. Accuracy of intraoral scans: an in vivo study of different scanning devices[J]. J Prosthet Dent, 2022,128(6):1303-1309. DOI: 10.1016/j.prosdent.2021.03.007 .
返回引文位置Google Scholar
百度学术
万方数据
[15]
Ke Y , Zhang Y , Wang Y ,et al. Comparing the accuracy of full-arch implant impressions using the conventional technique and digital scans with and without prefabricated landmarks in the mandible: an in vitro study[J]. J Dent, 2023,135:104561. DOI: 10.1016/j.jdent.2023.104561 .
返回引文位置Google Scholar
百度学术
万方数据
[16]
柯怡芳,张耀鹏,陈俊锴,. 几何特征辅助口内三维扫描仪获取牙列缺失种植修复数字印模的精度研究[J]. 中华口腔医学杂志, 2022,57(2):161-166. DOI: 10.3760/cma.j.cn112144-20210519-00258 .
返回引文位置Google Scholar
百度学术
万方数据
Ke YF , Zhang YP , Chen JK ,et al. Accuracy analysis of full-arch implant digital impressions when using a geometric feature[J]. Chin J Stomatol, 2022,57(2):161-166. DOI: 10.3760/cma.j.cn112144-20210519-00258 .
Goto CitationGoogle Scholar
Baidu Scholar
Wanfang Data
[17]
Li Y , Fang H , Yan Y ,et al. Accuracy of intraoral scanning using modified scan bodies for complete arch implant-supported fixed prostheses[J]. J Prosthet Dent, 2024,132(5):994.e1-994.e8. DOI: 10.1016/j.prosdent.2024.07.015 .
返回引文位置Google Scholar
百度学术
万方数据
[18]
Fu XJ , Liu M , Shi JY ,et al. Comparison of different intraoral scanners with prefabricated aid on accuracy and framework passive fit of digital complete-arch implant impression: an in vitro study[J]. Clin Oral Implants Res, 2025,36(1):17-27. DOI: 10.1111/clr.14353 .
返回引文位置Google Scholar
百度学术
万方数据
[19]
Mizumoto RM , Yilmaz B , McGlumphy EA Jr,et al. Accuracy of different digital scanning techniques and scan bodies for complete-arch implant-supported prostheses[J]. J Prosthet Dent, 2020,123(1):96-104. DOI: 10.1016/j.prosdent.2019.01.003 .
返回引文位置Google Scholar
百度学术
万方数据
[20]
Mizumoto RM , Yilmaz B . Intraoral scan bodies in implant dentistry: a systematic review[J]. J Prosthet Dent, 2018,120(3):343-352. DOI: 10.1016/j.prosdent.2017.10.029 .
返回引文位置Google Scholar
百度学术
万方数据
[21]
Diker E , Terzioglu H , Gouveia D ,et al. Effect of material type, torque value, and sterilization on linear displacements of a scan body: an in vitro study[J]. Clin Implant Dent Relat Res, 2023,25(2):419-425. DOI: 10.1111/cid.13187 .
返回引文位置Google Scholar
百度学术
万方数据
[22]
Mizumoto RM , Alp G , Özcan M ,et al. The effect of scanning the palate and scan body position on the accuracy of complete-arch implant scans[J]. Clin Implant Dent Relat Res, 2019,21(5):987-994. DOI: 10.1111/cid.12821 .
返回引文位置Google Scholar
百度学术
万方数据
[23]
Gómez-Polo M , Ballesteros J , Perales-Padilla P ,et al. Guided implant scanning: a procedure for improving the accuracy of implant-supported complete-arch fixed dental prostheses[J]. J Prosthet Dent, 2020,124(2):135-139. DOI: 10.1016/j.prosdent.2019.09.022 .
返回引文位置Google Scholar
百度学术
万方数据
[24]
Carneiro Pereira AL , Carvalho Porto de Freitas RF , de Fátima Trindade Pinto Campos M ,et al. Trueness of a device for intraoral scanning to capture the angle and distance between implants in edentulous mandibular arches[J]. J Prosthet Dent, 2022,128(6):1310-1317. DOI: 10.1016/j.prosdent.2021.02.039 .
返回引文位置Google Scholar
百度学术
万方数据
[25]
Pan Y , Tsoi J , Lam WY ,et al. Improving intraoral implant scanning with a novel auxiliary device: an in-vitro study[J]. Clin Oral Implants Res, 2021,32(12):1466-1473. DOI: 10.1111/clr.13847 .
返回引文位置Google Scholar
百度学术
万方数据
[26]
Brånemark PI . Osseointegration and its experimental background[J]. J Prosthet Dent, 1983,50(3):399-410. DOI: 10.1016/s0022-3913(83)80101-2 .
返回引文位置Google Scholar
百度学术
万方数据
[27]
Jemt T . Failures and complications in 391 consecutively inserted fixed prostheses supported by Brånemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup[J]. Int J Oral Maxillofac Implants, 1991,6(3):270-276.
返回引文位置Google Scholar
百度学术
万方数据
备注信息
A
赖红昌,Email: mocdef.6ab21ialgnahcgnoh,电话:021-53315299
B

富晓娇:伞形扫描杆的设计、研究的设计和执行、文章撰写;蔡征真:研究执行、数据分析、史俊宇、赖红昌:研究设计、文章修改

C
富晓娇, 蔡征真, 史俊宇, 等. 两种扫描杆辅助下全牙弓种植固定修复口内扫描精度的比较[J]. 中华口腔医学杂志, 2025, 60(3): 267-272. DOI: 10.3760/cma.j.cn112144-20240817-00315.
D
所有作者声明不存在利益冲突
E
中央高校基本科研业务费专项资金 (YG2024QNA24)
评论 (0条)
注册
登录
时间排序
暂无评论,发表第一条评论抢沙发
MedAI助手(体验版)
文档即答
智问智答
机器翻译
回答内容由人工智能生成,我社无法保证其准确性和完整性,该生成内容不代表我们的态度或观点,仅供参考。
生成快照
文献快照

你好,我可以帮助您更好的了解本文,请向我提问您关注的问题。

0/2000

《中华医学会杂志社用户协议》 | 《隐私政策》

《SparkDesk 用户协议》 | 《SparkDesk 隐私政策》

网信算备340104764864601230055号 | 网信算备340104726288401230013号

技术支持:

历史对话
本文全部
还没有聊天记录
设置
模式
纯净模式沉浸模式
字号